24
Guidance and manufacturer’s declaration – electromagnetic Emissions
The SUPERVEGA EVO SUCTION UNIT is intended for use in the electromagnetic environment specified below. The
customers or the user of the SUPERVEGA EVO SUCTION UNIT should make sure that it’s used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
Irradiated / Conducted
emissions CISPR11
Group 1
The SUPERVEGA EVO SUCTION UNIT only used RF energy only
for its internal functioning. Therefore, its RF emissions are very low
and are not cause interference in proximity of any Electronic appli-
ances.
Irradiated / Conducted
emissions CISPR11
Class [B]
The SUPERVEGA EVO SUCTION UNIT can be used in all environ-
ments, including domestic and those connected directly to the public
mains distribution that supplies power to environments used for do-
mestic scopes.
Harmonic emissions
EN 61000-3-2
Class [A]
Voltage fluctuations / flicker
emissions EN 61000-3-3
Complies
ENGLISH
Guidance and manufacturer’s declaration – Immunity Emissions
The SUPERVEGA EVO SUCTION UNIT is intended for use in the electromagnetic environment specified below. The
customers or the user of the SUPERVEGA EVO SUCTION UNIT should make sure that it’s used in such an environment.
Immunity Test
Level indicated by the
EN 60601-1-2
Compliance Level
Electromagnetic environment - guidance
Electrostatic
discharge (ESD)
EN 61000-4-2
+/-8kV on contact
+/-15kV in air
The device doesn’t
change its state
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic ma-
terial, the relative humidity should be at least
30%.
Electrical fast
transient / burst
EN 61000-4-4
± 2kV power supply
lines
± 1kV for input /
output lines
The device doesn’t
change its state
Mains power quality should be that of a typical
commercial environment or hospital.
Surge
EN 61000-4-5
± 1kV differential mode
+/-2 kV ordinary mode
The device doesn’t
change its state
Mains power quality should be that of a typical
commercial environment or hospital.
Loss of voltage, brief
voltage interruptions
and variations
EN 61000-4-11
<5% U
T
(>95% dip U
T
)
for 0,5 cycle
40% U
T
(60% dip U
T
)
for 5 cycle
70% U
T
(30% dip U
T
)
for 25 cycle
<5% U
T
(>95% dip U
T
)
for 5 sec
- -
Mains power quality should be that of a typi-
cal commercial environment or hospital If the
user of the SUPERVEGA EVO SUCTION
UNIT request that the appliance operates
continuously, the use of a continuity unit is
recommended.
Magnetic field with
network frequency
(50/60 HZ)
EN 61000-4-8
30A/m
The device doesn’t
change its state
The power frequency magnetic field should
be measured in the intended installation lo-
cation to make sure that it’s sufficiently low.
Note U
T
is the value of the power supply voltage