19
If the appliance still does not nebulize after having checked the above mentioned conditions,
it is recommended to contact the reseller or Gima’s technical service.
BEFORE CARRYING OUT ANY CHECKS IN THE EVENT OF ANOMALIES OR
MALFUNCTIONS, CONTACT GIMA’S TECHNICAL SERVICE. THE MANUFACTURER
OFFERS NO GUARANTEE FOR THE APPLIANCES WHICH HAVE ASSESSED AS TAM-
PERED AFTER THE CHECKS CARRIED OUT BY THE TECHNICAL SERVICE.
CLEANING AND WASHING OF ACCESSORIES
Before each use and / or after the cleaning operations, check the integrity of all accessories
supplied with the device. Turn off the device before each cleaning operation and disconnect
the power cable from the socket.
PREPARATION
1. Pull out the air hose from the nebulizer leaving it plugged to the air outlet pipe of the de-
vice.
2. Rotate the upper part of the nebulizer anticlockwise.
3. Disconnect the internal pisper at the bottom of the nebulizer with finger force.
CLEANING
Clean all the components of the nebulizer (except the air hose) before and after each use by
choosing one of the methods described below.
Type of defect
Cause
Corrective action
1. The device
doesn’t work.
The plug may be mis-
placed in the wall socket.
Thermal protector may be
on (the device has been
working beyond its limits
and / or near heat sources).
Make sure the plug is properly placed in
the wall socket. Make sure the ON/OFF
switch is in position I.
Switch off the device by pressing the
switch to position 0 and the motor cool
down for at least 30 minutes.
2. Poor atomization. Clogged ampoule.
Clean and disinfect the ampoule as set
out in the manual.
3. Poor atomization. Clogged ampoule.
If the washing had no positive outcome,
replace the ampoule.
4. Lack of
atomization.
Nozzle stuck.
Press hard the nozzle (cylindrical im-
mersion tube) placed inside the poly-
carbonate ampoule bottom with a finger.
5. Slow atomization. Excessive oily drug.
Dilute the drug with saline.
6. Noisy device.
Prolonged use.
Contact the reseller or Gima’s technical
service.
Defects 1 - 2 - 3 - 4
- 5 - 6
None of the corrective
actions has been effective.
Contact the reseller or Gima’s technical
service.