21

Electromagnetic compatibility (EMC)

The podoscope needs special precautions concerning electromagnetic 

compatibility (EMC) and must be installed and operated according to the 

EMC information contained in this manual.

In order to avoid possible risks of electromagnetic interferences, do not use 

radio frequence portable or mobile devices near the podoscope.

Generally the podoscope should not be used near or together other 

equipements; in the case this would not be possible, it’s necessary to observe 

its operativity to verify the normal functioning.

Emissions test

RF emissions

CISPR 11

RF emissions

CISPR 11

Harmonic emissions

IEC 61000-3-2

Voltage fluctuations/

flicker emissions

IEC 61000-3-3

Compliance

Group1

Class B

Compliant

Compliant

Electromagnetic environment – guidance

The equipment 27363 must emit electromagnetic 

energy in order to perform its intended function. 

Nearby electronic equipment may be affected.

The 27363 is suitable for use in all establishments 

other than domestic and those directly connected 

to the public low-voltage power supply network 

that supplies buildings used for domestic purposes

GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC EMISSIONS

The equipment 

27363

 is intended for use in the electromagnetic environment specified below. 

The customer or the user of the equipment 

27363

 should assure that it is used in such an environment

ENGLISH

120 cm 47.25 in

15 cm

5.9 in

Podoscope

Handle

Handle

100 cm

39.4 in

130 cm

51.19 in

Summary of Contents for 27363

Page 1: ...efore using the product AVIS Les op rateurs doivent lire et bien comprendre ce manuel avant d utiliser le produit ATENCI N Los operadores tienen que leer y entender completamente este manual antes de...

Page 2: ...ence norms 17 1 5 Electrical features and voltage 18 1 6 Components identification 18 Electrical schema 19 1 7 General safety warning 19 Chapter 2 INSTALLATION 20 2 1 Transport and stocking 20 2 2 Con...

Page 3: ...USAGE MAINTENANCE Instructions for a correct installation are contained in the par 2 2 CONDITIONS FOR THE INSTALLATION Follow the instructions for use Read carefully this manual for the installation...

Page 4: ...an discover variations of colour intensity on the foot plant corresponding to higher or lower plant pressure The base is of ABS A separable net cable with plug a jack and a led strip are its electric...

Page 5: ...sports centres by operators with knowledge of biomechanics like doctors in sport medicine orthopedic specialists and physiotherapists Reference harmonized norms CEI EN 60601 1 Safety of electromedical...

Page 6: ...assembled and fixed in such a solid way so as to stand up to the often adverse transport conditions As to indirect shocks the device is classified in Class II As to current leakage it can be classifi...

Page 7: ...always take the following precautions A Always unplug the podoscope during maintenance operations B Do not move the podoscope while using it C Each time before the usage check the functioning of the c...

Page 8: ...ystyrene polyethylene cardboard must follow your local waste disposal guidelines 3 Verify that the podoscope has not been damaged during transport In case please contact the manufacturer immediately I...

Page 9: ...00 3 2 Voltage fluctuations flicker emissions IEC 61000 3 3 Compliance Group1 Class B Compliant Compliant Electromagnetic environment guidance The equipment 27363 must emit electromagnetic energy in o...

Page 10: ...ment Mains power quality should be that of a typical commercial or hospital environment Mains power quality should be that of a typical commercial or hospital environment If the user of 27363 requires...

Page 11: ...the examination can be performed only if the patient is barefoot otherwise lay the caoutchouc on the transparent surface This material is also useful to check the validity of arch support treatments W...

Page 12: ...19 Then get out the blown fuse with a suitable instrument and replace it with a T 0 5 A 5x20 250V fuse If you have to replace the power cord the new one must have the same characteristics and certifi...

Page 13: ...he warranty GIMA will repair and or replace free of charge all the defected parts due to production reasons Labor costs and personnel traveling expenses and packaging not included All components subje...

Page 14: ...74...

Page 15: ...75...

Page 16: ......

Reviews: