MAINTENANCE

Care and cleaning

The probe is the most delicate part of the thermometer. Use with care when cleaning the probe
to avoid damage.

Delicately clean the lens with a cotton cloth (soft cloth) moistened with water or 

70% 

alcohol and allow the

probe to fully dry. Clean probe after each use to assure accurate reading and avoid crossed contamination.
Put on lid when not using the thermometer.

Battery replacement

The device is supplied with one lithium battery (CR2032 x 1); don’t forget to switch it off before  changing the
battery (see illustration). The positive (+) side of the battery must face up. Keep the battery away from
children.

Disposal: The product must not be disposed of along with other domestic waste. The users
must dispose of this equipment by bringing it  to a specific recycling point for electric and
electronic equipment.
For further information on recycling points contact the local authorities, the local recycling
center or the shop where the product was purchased. If the equipment is not disposed of
correctly, fines or penalties may be applied in accordance with the national legislation and
regulations.

GIMA WARRANTY CONDITIONS

Congratulations for purchasing a GIMA product.
This product meets high qualitative standards both as regards the material and the production. The warranty
is valid for 12 months from the date of supply of GIMA.
During the period of validity of the warranty, GIMA will repair and/or replace free of charge all the defected

Error
Message

Solution

Remove the battery, wait for 1 minute and
repower it. If the message reappears, contact
the retailer for service.

Wait until all the icons stops flashing.

Allow the thermometer to rest in a room for
at least 30 minutes at room temperature:
10°C and 40°C (50°F ~104°F).

Please select the target within specifications.
If a malfunction still exists, please contact
the nearest retailer.

Change with a new battery.

Problem

Error 5~9, the system is not functioning
properly.

Measurement before device stabilization.

The device shows a rapid ambient
temperature change.

The ambient temperature is not within the
range between 10°C and 40°C (50°F ~104°F).

Forehead: Temperature taken is higher
than +42.2°C (108°F). Scan: Temperature
taken is higher than +80°C (176°F)

Forehead: Temperature taken is lower
than +34°C (93.2°F). Scan: Temperature
taken is lower than -22°C (-7.6°F)

LCD display cannot be powered on to the
ready stage.

TROUBLESHOOTING

ENGLISH

Summary of Contents for 25585

Page 1: ...ION The operators must carefully read and completely understand the present manual before using the product AVIS Les opérateurs doivent lire et bien comprendre ce manuel avant d utiliser le produit ACHTUNG Die Bediener müssen vorher dieses Handbuch gelesen und verstanden haben bevor sie das Produkt benutzen ATENCIÓN Los operadores tienen que leer y entender completamente este manual antes de utili...

Page 2: ...errange 4 or 2 C 4 F whicheverisgreater Battery one lithium battery CR2032 FUNCTION Before each use make sure the probe is clean and without damage Press ON MEM button The thermometer is ready for use after you hear two beep sounds Setting date and time Correctly setting the clock will assure a proper interpretation of the stored measurements Press and hold the ON MEM button for 5 seconds When the...

Page 3: ...ture the thermometer should be placed on a table far from elements that can affect the measurement Wait approx 15 minutes Memorizing temperatures The memory function can be used to show the last 9 measurements with date time temperature and mode Press the ON MEM button when power is on to see the temperatures stored Stop watch After switching on the thermometer press and hold the ON MEM button and...

Page 4: ...slation and regulations GIMA WARRANTY CONDITIONS Congratulations for purchasing a GIMA product This product meets high qualitative standards both as regards the material and the production The warranty is valid for 12 months from the date of supply of GIMA During the period of validity of the warranty GIMA will repair and or replace free of charge all the defected Error Message Solution Remove the...

Page 5: ...d personnel or with non original spare parts defects caused by negligence or incorrect use GIMA cannot be held responsible for malfunctioning on electronic devices or software due to outside agents such as voltage changes electro magnetic fields radio interferences etc The warranty is void if the above regulations are not observed and if the serial code if available has been removed cancelled or c...

Page 6: ...36 ...

Reviews: