background image

8

Fever alarm

If thermometer detects a body temperature greater than 37.5° C there will be a long beep
sound followed by three short beep sounds.

Temperature level colored LED indicator

If the thermometer measures a body temperature lower than  37.5 ° C, the led will be green,
If the temperature is greater than  37.5 ° C and 38.0 ° C, the led will be yellow, if the
temperature is > = 38.0 ° C, the led will  be red.

Measuring temperature under Scan mode

This mode is not suitable for the measurement of human body temperature.

Scan mode measures the temperature of objects and liquids (for example milk or bath water). After switching
on the device, press and hold the ON/MEM button, and press SCAN once until the LCD display shows the
icon 

 for “infrared thermometer” mode. In this mode it is possible to measure the surface temperature of

objects and liquids.
If you press and hold the “SCAN” button, the reading of measurement will be continuously updated.

After measurement

Device will automatically shut off if left idle for more than 1 minute to extend battery life. The
LCD will only show the date, time and room temperature   in stand-by mode. For  an
accurate measurement of room temperature, the thermometer should be placed on a table,
far from elements that can affect the measurement. Wait approx. 15 minutes.

Memorizing temperatures

The memory function 

  can be used to show the last 9 measurements with date, time,

temperature and mode. Press the ON/MEM button when power is on to see the temperatures
stored.

Stop watch

After switching on the thermometer, press and hold the ON/MEM button and press the
SCAN button twice for stop watch mode to see the icon appear. In this mode you can count
by 0.01 seconds to three minutes. To start or stop the count press the SCAN button. To set
count to zero press the ON/MEM button. To exit function press and hold the ON/MEM
button and press the SCAN button.

WARNINGS

- Temperature can be influenced by several external factors; variation between one measurement
and the other are normal.
- Before the measurement, please stay in a stable environment for 5 mins.
- Avoid bath or exercise before measuring body temperature;  keep the measurement area
clean and dry, free from cosmetics or creams
- During measurement avoid direct exposure to sun, rain or wind; do not measure near heat
sources or in excessively humid places.
- Always keep device at a temperature between -20 and 50°C (-4 and122°F), in a dry place
away from liquids and direct sunlight.
- in case of extraordinary maintenance interventions, contact the nearest retailer.

ENGLISH

Summary of Contents for 25585

Page 1: ...ION The operators must carefully read and completely understand the present manual before using the product AVIS Les opérateurs doivent lire et bien comprendre ce manuel avant d utiliser le produit ACHTUNG Die Bediener müssen vorher dieses Handbuch gelesen und verstanden haben bevor sie das Produkt benutzen ATENCIÓN Los operadores tienen que leer y entender completamente este manual antes de utili...

Page 2: ...errange 4 or 2 C 4 F whicheverisgreater Battery one lithium battery CR2032 FUNCTION Before each use make sure the probe is clean and without damage Press ON MEM button The thermometer is ready for use after you hear two beep sounds Setting date and time Correctly setting the clock will assure a proper interpretation of the stored measurements Press and hold the ON MEM button for 5 seconds When the...

Page 3: ...ture the thermometer should be placed on a table far from elements that can affect the measurement Wait approx 15 minutes Memorizing temperatures The memory function can be used to show the last 9 measurements with date time temperature and mode Press the ON MEM button when power is on to see the temperatures stored Stop watch After switching on the thermometer press and hold the ON MEM button and...

Page 4: ...slation and regulations GIMA WARRANTY CONDITIONS Congratulations for purchasing a GIMA product This product meets high qualitative standards both as regards the material and the production The warranty is valid for 12 months from the date of supply of GIMA During the period of validity of the warranty GIMA will repair and or replace free of charge all the defected Error Message Solution Remove the...

Page 5: ...d personnel or with non original spare parts defects caused by negligence or incorrect use GIMA cannot be held responsible for malfunctioning on electronic devices or software due to outside agents such as voltage changes electro magnetic fields radio interferences etc The warranty is void if the above regulations are not observed and if the serial code if available has been removed cancelled or c...

Page 6: ...36 ...