Parts Replacement
deUTeRiUM LAMp (d2)
Resets the deuterium lamp age time and sets a new index of time counter.
TUnGSTen LAMp (W)
Resets the tungsten lamp age time.
Calibrations
ABSORBAnCe
Using this item, the absorption response can be slightly corrected. The absorbance coefficient
is the multiplier of the basic calculated value.
inTenSiTY
This item is used to automatically calibrate the spectral intensity to 100%. The instrument must be in
measuring mode (lamps on). calibration takes about 12 seconds and the service menu is exited, when
finished. For more information, refer to
Intensity Calibration after Lamp(s) Replacement on page 28
WAVeLenGTHS
Use this item to manually adjust the wavelengths settings (scale and offset).
NOTE
The setting is related to the mechanical setting of the monochromator.
Factory Settings
pARAMeTeRS ReSeT
Use this item to set all device parameters to default values.
CALiBRATiOnS ReSeT
Use this item to set device calibrations to default values, individually for absorbance, intensity,
and wavelengths.
OTHeRS ReSeT
Use this item to reset other device records, individually for web password, errors log, and action
log.