Diagnostics of the actual Io InTERFAcE status are accessed by pressing the
MenU
key and then
selecting
diagnostics
|
interface
. For digital inputs value oPEn (H) means opened input, high
level and cloSE (l) means closed input, low level. For digital output is displayed oFF (for open
switch) and on (for closed switch).
The analog input, digital input 1 and 2, and digital output switches can be read through remote
control (ETHERnET).
diGiTAL inpUT (din1 And din2)
Function setting for inputs DIn1 and DIn2 on external interface:
SETTING
DESCRIPTION
OFF
No function.
MAKE AUTOZERO
Pulse L causes the detector to be reset – autozeroed (AUTOZERO).
LAMP INIT
Pulse L causes initialization of the light sources (LAMP ON).
LAMP DEINIT
Pulse L causes deinitialization of the light sources (LAMP OFF).