51
FRonT PAnEl conTRol
| USER’S GUIDE
F
R
O
n
T
p
A
n
e
L
C
O
n
TR
O
L
F
R
O
n
T
p
A
n
e
L
C
O
n
TR
O
L
deUTeRiUM LAMp (d2)
The following values are displayed:
Current Values of D2 (Deuterium Lamp) Control
NAME
DESCRIPTION
VA3
Voltage at the output of the current regulator behind the measuring resistor (approx. +72 V).
VA4
Voltage at the output of the current regulator in front of the measuring resistor (approx. +82 V).
VA5
Voltage at the anode of the lamp (approx. +71 V).
IAD
Anode lamp current (approx. +0.282 A).
VHT
Heating voltage in steady state (approx. +1 V).
IHT
Heating current (approx. +2 A).
TUnGSTen LAMp (W)
The following values are displayed:
Current Values of W (Tungsten Lamp) Control
NAME
DESCRIPTION
VLO
Current regulator Lo voltage (approx. +0.9 V).
VHI
Current regulator Hi voltage (approx. +11.7 V).
VBL
Lamp voltage (approx. +10.8 V).
IBL
Lamp current (approx. +0.8 A).
Cooling
Shows temperature of illuminator TEMP - If the cooling sensor is not connected, the display will
show
TeMp: SenSOR nOT inSTALLed
, rotation speed of fans FAn1 and FAn2 (rpm) and their
voltage (V).
Interface
Shows the status of the analog and digital Io InTERFAcE:
Status of digital inputs DIn1 and DIn2 and digital output DoUT1 and DoUT2.
Summary of Contents for VERITY 1741
Page 1: ...User s Guide VERITY 1741 UV VIS Detector...
Page 2: ......
Page 4: ......
Page 10: ...This page intentionally left blank...
Page 24: ...This page intentionally left blank...
Page 28: ...This page intentionally left blank...