20
To fit the lower part on the body,
Press on the push-button.
Insert the lower part into
the ejection rod. Release
the push-button.
Screw the connecting nut
turning clockwise.
Press on the tip-ejector
button.
Insert the ejector clip and
release the tip-ejector
button.
chemical decontamination
you may choose to decontaminate your pipette chemi-
cally, in accordance with your own procedures. Whatever
decontaminant you use, check with the supplier of the
decontaminant that it is compatible with stainless steel
and the plastics used in the construction of the pipette:
PA (Polyamide), PBT (Polybutylene Terephtalate), PC
(Polycarbonate), PC/PBT (Polycarbonate/ Polybutylene
Terephtalate), POM (Polyoxymethylene), or PVDF (Poly-
vinylidene Fluoride).
Single and Multichannel - Upper Part (handle)
Wipe the upper part (handle) of the pipette with a
soft-cloth or lint-free tissue impregnated with the
chosen decontaminant.
Wipe the upper part of the pipette with a soft- cloth
or lint-free tissue soaked with distilled water or
sterile water.
Single channel - Lower Part (Volumetric module)
The following components
only
can be immersed in a
decontaminant solution: connecting nut, tip-ejector,
tip-holder, piston assembly, seal and O-ring.