7
3) Select a piston and slide the stem
2
fully into the
clamp.
4) Slide the mounted piston into the capillary
4
.
5) Gently push the capillary until it snaps onto the
capillary-holder
6
.
cP10, cP25, cP50, cP100, cP250
and cP1000 tipack
The capillary-pistons (CP) are deliv-
ered already assembled. To fit a cap-
illary-piston, follow this procedure:
1) Press the push-button on the
MiCroMan past the first stop
position to the second stop po-
sition. The piston clamp should
now be open and protruding
from the capillary-holder.
2) Slide the MiCroMan piston
clamp inside the top of the capil-
lary tube until the piston clamp is
halfway down the piston mount-
ing stem.
3) Slowly release the push-button,
while pushing the body of the
MiCroMan into the capillary-
piston until it is firmly seated on
the MiCroMan.
4) Continue releasing the push-
button to ensure that the piston
is secured by the piston clamp.
5) To ensure that the piston is
correctly seated, and therefore
calibrated, slowly press the push-
button until you feel and hear a
slight click.
Step 2
Step 3
1
2
3
4
CP