Pipetting

Place a vessel containing the solution to be pipetted in the front location of the right 

drawer. The solution should be under the pipetting head.

Liquids containing protein solutions and organic solvents can leave a film of liquid 

on the inside wall of the tip; pre-rinse the tip to minimize any errors related to 

this phenomenon. Pre-rinsing consists of aspirating the first volume of liquid and 

then dispensing it back into the same vessel (or into a waste basket). Subsequent 

volumes pipetted will have levels of accuracy and precision within specifications. As 

recommended by the ISO 8655 standard, pre-rinse to stabilize dead volume inside 

the pipetting head

ASPIRATION

• 

Gently squeeze the pipetting handle

       

to the first stop

 (this corresponds to the 

set volume of liquid).

• 

Use the pipetting handle        to lower 
the pipetting head

       

to immerse the 

tips in the liquid (see immersion depth 
table below). 

It is important that the 

tips do not touch the bottoms of the 

wells.

• 

Slowly retract the pipetting handle to 

aspirate the liquid into the tips.

• 

Move the pipetting head fully up.

To simplify this operation, the height 
adjuster

       

can be set by turning the 

upper screw so that the pipetting head 
may only be moved down to a position 
slightly above the bottom.

MODEL

IMMERSION DEPTH 

(MILLIMETERS)

WAIT TIME 

(SECONDS)

P20

2-3

1

P220

2-4

1

Check the tip ejector when fitting tips. If the tips have shifted this 

implies that they have not been fitted correctly.

The maximum 

immersion depth 

of the tips can be 

adjusted using a 

height adjuster.

1

2

3

1

1

2

3

NOTE

NOTE

O

PE
R

A

TIO

N

9

PLATEMASTER®

 

|

 

USER’S GUIDE

Summary of Contents for F110761

Page 1: ...User s Guide EN...

Page 2: ...iption 7 Moving PLATEMASTER 7 Height Adjuster 7 Volume Setting 8 Tip Fitting 8 Pipetting 9 Pipetting 384 wells 10 Tip Ejection 11 General Guidelines for Good Pipetting 12 ACCESSORIES 13 ASSOCIATED TIP...

Page 3: ...pipetting steps Figure 1 Gilson PLATEMASTER P20 PLATEMASTER is suitable for a wide range of applications Pipetting and providing master mixes for PCR qPCR Simultaneous processing of nearly all reacti...

Page 4: ...filter tips installed can be autoclaved Gilson is not liable for any damage resulting from the use of PLATEMASTER Follow the instructions in this user s guide to ensure durability and top performance...

Page 5: ...vided by the chemical manufacturer before cleaning up the spill and take all safety precautions required Remove PLATEMASTER pipetting head and PLATEMASTER base from box Insert the pipetting head as fo...

Page 6: ...F110761 F110762 Volume Range 0 5 to 20 L Volume can be adjusted in 0 1 L increments 2 to 220 L Volume can be adjusted in 1 L increments Temperature Range Recommended working temperature 20 C to 25 C...

Page 7: ...ee height adjusters positions Moving PLATEMASTER Whether placed on the lab bench under a fume hood in a cold room or used in the field PLATEMASTER is completely portable PLATEMASTER includes a locking...

Page 8: ...le Move the pipetting head to the left side Slide the right drawer out and place a new tip rack on the back position The four locations have been specially designed for PIPETMAN DIAMOND Tips TIPACK Sl...

Page 9: ...ze the pipetting handle to the first stop this corresponds to the set volume of liquid Use the pipetting handle to lower the pipetting head to immerse the tips in the liquid see immersion depth table...

Page 10: ...pper screw so that the pipetting head may only be moved down to a position slightly above the bottom Pipetting 384 well Microplates PLATEMASTER is able to load 384 well plates with only four movements...

Page 11: ...Move the pipetting head over the microplate and dispense the last 96 wells The 384 well microplate is loaded in less than one minute requiring only four liquid aspiration steps Tip Ejection Move the...

Page 12: ...id used Never allow any liquid to enter the pin plate see Figure 14 page 15 This phenomenon can be prevented by squeezing and retracting the pipetting handle slowly and gently Change the tips before a...

Page 13: ...ad height adjuster F1077605 Lubrication box for O rings PLATEMASTER F1077606 AmpliPur Expert Tips Pin plate Assembly for PLATEMASTER F110765 AmpliPur Expert Tips Reload Block pack of 2 for PLATEMASTER...

Page 14: ...0ST Validated for use with 384 well plates PIPETMAN DIAMOND Tips DL10 DL10ST DSL10 and DSL10ST could be used under a specific product configuration For more information please contact Gilson Technical...

Page 15: ...EMASTER to excessive UV radiation and never overnight Do not autoclave any part of PLATEMASTER Material in Contact with Liquids or Vapors FKM Fluoroelastomers EPDM ethylene propylene diene monomer PP...

Page 16: ...PIPETMAN DIAMOND Tips D300 and AmpliPur Expert Tips Each PLATEMASTER is inspected and validated by qualified technicians in accordance with the Gilson Quality System The adjustment is carried out und...

Page 17: ...in the event devices are altered or modified by any party other than Gilson or its designates Gilson s sole liability under this warranty shall be limited to at Gilson s sole option repair or replacem...

Page 18: ...NOTES Notes 18 Notes PLATEMASTER...

Page 19: ...NOTES Notes 19 PLATEMASTER USER S GUIDE...

Page 20: ...e without notification errors excepted Reproduction adaptation or translation without prior written permission is prohibited except as allowed under copyright laws LT801549 I 2022 Gilson Inc All right...

Reviews: