DISPENSING
•
Place a 96-well microplate at the front location of the left drawer
and keep it
fully retracted in the rear position.
•
Move the pipetting head
to the left and align the tips with the microplate.
•
Pull the pipetting head slowly down into
the wells, but do not touch the bottom
of the wells.
•
Squeeze the pipetting handle
slowly to dispense all the solution.
•
Pull the pipetting head slowly up with
the pipetting handle fully squeezed.
•
Gently retract the pipetting handle.
•
To simplify this operation, a height
adjuster
can be set by turning the
upper screw so that the pipetting head
may only be moved down to a position
slightly above the bottom.
Pipetting 384-well Microplates
PLATEMASTER is able to load 384-well plates with only four movements using the
384-well plate adapter.
STEP 1:
• Place a 384-well microplate on the 384-well plate adapter and make sure that it
is located at the first step by checking the position on the blue wheel.
• Aspirate the solution.
• Move the pipetting head over the microplate and dispense the first 96 wells.
STEP 2:
• Move the microplate on to the second position of the 384-well plate adapter
using the blue wheel.
• Aspirate the solution. Move the pipetting head over the microplate and
dispense the next 96 wells.
1
2
3
4
4
3
2
1
O
PE
R
A
TIO
N
10
OPERATION
|
PLATEMASTER®