background image

Gilson Aqua-Check Moisture Tester, MA-26X 

9

For help with operation or technical issues, 

Email: [email protected] or Call: 800.444.1508

5.0  TROUBLESHOOTING

 

Low 

High

  Readings 

Readings

  Possible Cause

Solution

GILSON AQUA-CHECK TROUBLESHOOTING

 

Testing procedure errors 

Review instructions and procedures.

 

Incorrect sample weights 

Check balance operation and re-zero.

 

Errors in corrected values 

Change in material type. Repeat correlation procedure with new material.

 

 

Insufficient agitation 

Some cohesive materials may require 3 minutes or more. Use Steel Pulverizing Balls.

 

 

Insufficient quantity of reagent 

Add an additional scoop of reagent.

 

 

Ineffective reagent 

Replace with fresh MAA-44 Reagent.

 

 

Premature contact between reagent 

Ensure vessel is held horizontal when mating with cap. 

 

 

 

and specimen 

 

 

Insufficient sample preparation 

Break cohesive material into finer sizes and use Pulverizing Balls.

 

 

Inadequate pressure seal 

Check and clean chamber cap gasket. Inspect Vessel, cap, and stirrup for damage  

 

 

 

 

or cracks. Replace gasket if necessary.

 

Defective gauge 

Gauge needle should rest at zero and move smoothly. Replace if defective.

 

 

Low temperature 

Run two or three test cycles to warm up Vessel.

 

 

High temperature 

Allow Vessel to cool off.

 

 

Moisture inside vessel prior to testing 

Insure Vessel and cap are completely dry before testing.

 

Gauge reading error 

Hold vessel horizontally at eye level when reading gauge.

 

 

Residue from previous tests 

Use included Coarse-Bristle Brush to clean vessel thoroughly between tests.

MAA-47

MAA-48

6.0 ACCESSORIES

Calcium Carbide Reagent, 10lb can¹ 

MAA-44

Aqua-Check 0–20% Pressure Gauge 

MAA-45

Electronic Balance, 220x0.1g 

OB-205

1.25in (32mm) Steel Balls, pkg. 2 

MAA-47

Long-Handle Reagent Scoop 

MAA-48

Large, Coarse-Bristle Brush 

MAA-51

Small, Fine-Bristle Brush 

MAA-50

Sample Cup 

MAA-52

Heavy-Duty Waterproof Plastic Case 

MAA-46

¹ Due to shipping restrictions, reagent is sold separately in 10lb (4.5kg) cans.  

Material from these larger cans can be used to replenish smaller containers for 

field use. 

GILSON AQUA-CHECK ACCESSORIES

Description

Model

Summary of Contents for Aqua-Check MA-26X

Page 1: ...isture Tester OPERATING MANUAL Rev 5 2022 PHONE 800 444 1508 P O Box 200 Lewis Center Ohio 43035 0200 740 548 7298 Email customerservice gilsonco com Product Web Page globalgilson com FAX 800 255 5314...

Page 2: ...k Moisture Tester MA 26X 2 Safety Instructions 3 1 0 Unpacking Set Up 4 2 0 Introduction 4 3 0 Operating Instructions 5 4 0 Accuracy 8 5 0 Troubleshooting 9 6 0 Accessories 9 Gilson Aqua Check Moistur...

Page 3: ...t procedures WARNING The calcium carbide reagent used with this test must be handled with great care When exposed to water calcium carbide produces highly flammable or explosive acetylene gas Testing...

Page 4: ...o order are listed separately in this manual Calcium Carbide Reagent is available separately from Gilson as MAA 44 in a 10lb can 2 0 INTRODUCTION The Gilson Aqua Check Moisture Tester is designed to m...

Page 5: ...il the balance shows the required 20g NOTE If the moisture content of the material is expected to exceed 20 a 10g sample should be used but the value read on the gauge must then be doubled It is also...

Page 6: ...mber STEP 6 Holding the Aqua Check Vessel horizontally place the chamber cap on the end of the chamber Position the black dot on the cap so that it will be facing away from your body when the pressure...

Page 7: ...uld be repeated for 1 3 minutes or until the gauge reading in Step 8 has stabilized b With Pulverizing Balls Hold the Aqua Check horizontally and use an orbital motion to spin the balls around inside...

Page 8: ...ressure Tester Method be followed This will ensure accurate and reliable results For optimum accuracy with most materials the Aqua Check results should be correlated with oven dry tests for each mater...

Page 9: ...and clean chamber cap gasket Inspect Vessel cap and stirrup for damage or cracks Replace gasket if necessary X X Defective gauge Gauge needle should rest at zero and move smoothly Replace if defective...

Reviews: