Gilson Aqua-Check Moisture Tester, MA-26X 

4

 1.0  UNPACKING & SET UP

NOTE:

 Each Gilson Aqua-Check is carefully inspected prior to packaging and shipment. Upon 

receipt of the shipment, immediately unpack and inspect the apparatus for signs of damage. 
Report damage immediately to the shipping company. Retain all shipping records and packing 
materials until proper operation is confirmed.

The complete MA-26 Aqua-Check Kit consists of the following components:

•  Cast Aluminum Sample Chamber Vessel with integral 0—20% Pressure Gauge
•  200x0.1g Electronic Balance for weighing of test specimens
•  Two 1.25in (32mm) Steel Pulverizing Balls
•  Long-Handled Reagent Scoop
•  Large, Coarse-Bristle Brush
•  Small, Fine-Bristle Brush
•  Sample Cup
•  Heavy-Duty plastic Case 

The MA-26X includes all of the components listed, except the Calcium Carbide Reagent*. Any components 
or  accessories  are  available  individually  from  Gilson.  Model  numbers  to  order  are  listed  separately  in 
this manual.

*Calcium Carbide Reagent is available separately from Gilson as MAA-44 in a 10lb can.

 2.0  INTRODUCTION

The  Gilson  Aqua-Check  Moisture  Tester  is  designed  to  measure  the  water  (moisture)  content  of  soils, 
aggregates,  and  other  non-reactive  materials.  The  Aqua-Check  apparatus  consists  of  an  aluminum 
sample chamber vessel with integral pressure gauge, electronic balance for sample weighing, two steel 
pulverizing balls, and assorted brushes and containers as required for proper completion of the test. 

Pulverized calcium carbide is used as a reagent, and reacts when combined with available water in the 
sample material to form acetylene gas. The increase of gas pressure in the sealed sample chamber is 
proportional to the water content, and is shown on the pressure gauge. The most accurate results are 
obtained when pressure gauge readings are correlated with data from oven-dry methods. For example, 
when testing soils for example, ASTM Standard Test Method D4944 provides a procedure to correlate 
Aqua-Check results directly with laboratory oven drying methods.

MA-26X

Summary of Contents for Aqua-Check MA-26X

Page 1: ...isture Tester OPERATING MANUAL Rev 5 2022 PHONE 800 444 1508 P O Box 200 Lewis Center Ohio 43035 0200 740 548 7298 Email customerservice gilsonco com Product Web Page globalgilson com FAX 800 255 5314...

Page 2: ...k Moisture Tester MA 26X 2 Safety Instructions 3 1 0 Unpacking Set Up 4 2 0 Introduction 4 3 0 Operating Instructions 5 4 0 Accuracy 8 5 0 Troubleshooting 9 6 0 Accessories 9 Gilson Aqua Check Moistur...

Page 3: ...t procedures WARNING The calcium carbide reagent used with this test must be handled with great care When exposed to water calcium carbide produces highly flammable or explosive acetylene gas Testing...

Page 4: ...o order are listed separately in this manual Calcium Carbide Reagent is available separately from Gilson as MAA 44 in a 10lb can 2 0 INTRODUCTION The Gilson Aqua Check Moisture Tester is designed to m...

Page 5: ...il the balance shows the required 20g NOTE If the moisture content of the material is expected to exceed 20 a 10g sample should be used but the value read on the gauge must then be doubled It is also...

Page 6: ...mber STEP 6 Holding the Aqua Check Vessel horizontally place the chamber cap on the end of the chamber Position the black dot on the cap so that it will be facing away from your body when the pressure...

Page 7: ...uld be repeated for 1 3 minutes or until the gauge reading in Step 8 has stabilized b With Pulverizing Balls Hold the Aqua Check horizontally and use an orbital motion to spin the balls around inside...

Page 8: ...ressure Tester Method be followed This will ensure accurate and reliable results For optimum accuracy with most materials the Aqua Check results should be correlated with oven dry tests for each mater...

Page 9: ...and clean chamber cap gasket Inspect Vessel cap and stirrup for damage or cracks Replace gasket if necessary X X Defective gauge Gauge needle should rest at zero and move smoothly Replace if defective...

Reviews: