Gill Instruments Ltd
_____________________________________________________________________________________________________________
________________________________________________________________________________________________
WindObserver II (WindObserver FC)
Page 27
Issue 2
Doc. No. 1390-PS-0011
June 2007
9.
MAINTENANCE & FAULT-FINDING
9.1. Cleaning
If there is any build up of deposit on the unit, it should be gently cleaned with a cloth,
moistened with soft detergent. Solvents should not be used, and care should be taken to
avoid scratching any surfaces. The unit must be allowed to defrost naturally after being
exposed to snow or icy conditions, do NOT attempt to remove ice or snow with a tool.
Do NOT remove black “rubber” transducer caps.
9.2. Servicing
There are no moving parts or user-serviceable parts requiring routine maintenance.
Opening the unit or breaking the security seal will void the warranty and the calibration.
In the event of failure, prior to returning the unit to your authorised Gill distributor, it is
recommended that:
All cables and connectors are checked for continuity, bad contacts, corrosion etc.
A bench test is carried out as described in Section 9.6.
You contact your supplier for advice
9.3. Fault-finding
Symptom
Solution
No output
Check DC power to WindObserver II (WindObserver FC), cable
and connections.
Check comms settings of WindObserver II (WindObserver FC)
(as detailed in Section 8 and host system match, including correct
Com port
Check unit is in Continuous mode
Check that in-line communication devices are wired correctly.
NOTE: It is usual for Anemometer TX + to be connected to
converter device RX +
Corrupted output
Check comms settings of WindObserver II (WindObserver FC)
and host system match.
Try a slower baud rate.
Check cable lengths and type of cable.
One way
communication
Check wiring is in accordance with the manual.
Failed / Incorrect
WindObserver II
(WindObserver
FC) output, data
invalid flag
Check that transducer path is not blocked