Page 44
12 February, 2013
speed in Metres/Sec )
and a 2 digit Wind only
Status Code.
0D0!
Retrieve a line of the
above data.
a<dir><mag><status><CR><LF
>
0+029+000.01+00<CR><LF>
aM3!
Temperature (degrees
C), Humidity (%),
Dewpoint (Degrees C)
and Full Sensor Status
Codes.
atttn<CR><LF>
00054<CR><LF>
0D0!
Retrieve a line of the
above data.
a<temp><humidity><dewpoint>
< CR><LF
0+021.6+041.0+007.8+00
<CR><LF>
aM4!
Pressure (Hecto
Pascals/millibars).
atttn<CR><LF>
00051<CR><LF>
0D0!
Retrieve a line of the
above data.
a<pressure><CR><LF>
0+1013.0<CR><LF>
AM8!
Digital Input 1.
atttn<CR><LF>
00051<CR><LF>
0D0!
Retrieve a line of the
above data.
a<dig1><CR><LF>
0+0408.000<CR><LF>
aM9!
Power Voltage (Volts)
atttn<CR><LF>
00011<CR><LF>
0D0!
Retrieve a line of the
above data.
a<voltage><CR><LF>
0+12.5<CR><LF>
CRC Commands
Command
Description
Response
Example
aMC!
Wind Direction/Speed
(direction in Degrees,
speed in Metres/Sec ),
Temperature (degrees-
C), Pressure (Hecto
Pascals/millibars)
Humidity (%) and CRC.
atttn<CR><LF>
a is unit identifier.
ttt is time in seconds.
n is number of data values.
Measurement command to
retrieve a reading of the
maximum time the MetPak
RG will take to complete a
measurement, have data
ready and the number of
data values.
00056<CR><LF>
0 is unit identifier.
005 is 5 sec. measurement.
6 is the number of data readings
(Direction, Speed, Temperature,
Pressure, Humidity and CRC).
0D0!
Retrieve a line of the
above data.
a<dir><mag><temp><pressure>
<humidity><CRC><CR><LF>
0+139+000.06+022.2+1017.8
+042.7+00BIQ<CR><LF>
aMC1!
Wind Direction/Speed
(direction in Degrees,
speed in Metres/Sec ),
Temperature (degrees
C), Pressure (Hecto
Pascals/millibars),
Dewpoint (degrees C)
and CRC.
atttn<CR><LF>
00056<CR><LF>
0D0!
Retrieve a line of the
above data.
a<dir><mag><temp><pressure>
<dewpoint><CRC><CR><LF
0+140+000.06+022.2+1017.7
+009.0+00D{Z<CR><LF>