Gill Instruments Ltd
_____________________________________________________________________________________________________________
________________________________________________________________________________________________
WindObserver II Pipe Mount
Page 11
Issue 04
Doc. No. 1390-PS-0009
August 2009
Connecting to a PC with a RS422 or RS485 input
The PC requires either an internal RS422 interface card, or an external RS422/485 to
RS232 converter, which will plug into the standard 9 way or 25 way RS232 serial port of
the PC or host equipment.
WindObserver II (Pipe Mount)
6 Pair Cable
PC with RS422 or
RS485 input
Signal
names
Pair
no.
Colour
Cable – twisted
pairs
Signal names
TXB (+)
TXA (–)
1
1
Green
Black
RXB (+)
RXA (–)
RXB (+)
RXA (-)
DIG 0V
2
2
4
White
Black
Blue
TXB (+)
TXA (–)
Sig. Gnd
Connected Internally
Screen
wire
Ground (Earth)
Supply V+
Supply V-
3
3
Red
Black
+
-
DC Power
supply 9 – 30V
See Section 6.6
Heater -
5
5
Yellow & Brown
Black & Black
+
-
Heater power
supply
See Section 6.6
陕西威瑞仪器仪表有限公司 生产代理销售:实验检测设备气象科学仪器设备