Operation
Manual 21568, Rev. F, March 2008
2.3.6
Real Time Pulse Profile and Sample Position Display
The Detector Out connector on the rear of the 8035XA Series Peak Power Sensor can be
connected to any common oscilloscope for a real-time amplitude profile of your signal, delayed
by about 120 ns.
Connect the SMB to BNC cables to your oscilloscope (digital scope preferred - especially for
sample delay setting >500
µ
s) as shown in Figure 2-12.
Figure 2-12: Pulse Profile and Sample Delay Test Setup
Use the Sample Delay output as an oscilloscope trigger source. This waveform rises at the
trigger point and falls at the sample point; thus, it provides both a stable scope trigger source
and a precise indicator of the trigger point and sample point.
The time length of the sample delay pulse is the sum of the sample delay which is displayed on
the Series 8540 power meter front panel, and the sample delay offset which is available
through the menu.
Summary of Contents for 80350A
Page 4: ......
Page 10: ...Series 8035XA Peak Power Sensors vi Manual 21568 Rev F March 2008 ...
Page 13: ...Manual 21568 Rev F March 2008 ...
Page 16: ...Series 8035XA Peak Power Sensors xii Manual 21568 Rev F March 2008 ...
Page 17: ...Manual 21568 Rev F March 2008 ...
Page 18: ...Series 8035XA Peak Power Sensors xiv Manual 21568 Rev F March 2008 ...
Page 24: ...Series 8035XA Peak Power Sensors 1 6 Manual 21568 Rev F March 2008 ...
Page 90: ...Series 8035XA Peak Power Sensors 6 22 Manual 21568 Rev F March 2008 ...
Page 92: ......
Page 93: ......
Page 94: ......
Page 95: ......
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ......
Page 100: ......
Page 101: ......
Page 102: ......
Page 103: ......
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......