Giga-tronics 2500B Series Microwave Signal Generators
Performance Verification
Operation Manual, Part Number 34737, Rev C, June 30, 2011
Page 113
9.2.9
Amplitude Modulation Accuracy
The following procedure tests the amplitude modulation accuracy at frequencies within each band.
9.2.9.1
Equipment and Materials
Microwave spectrum analyzer
Function generator
Oscilloscope
Digital voltmeter
Cables and adapters for connecting the test setup
Datasheet 6 (Table 72 on page 115)
Figure 52: Amplitude Modulation Accuracy and Bandwidth Test Setup
NOTE:
Set the output amplitude (into 50 Ohms) of the function generator accurately using a precision digital voltmeter.
NOTE:
The external frequency reference (timebase) should be “Tee” connected from the 2500B to both the spectrum
analyzer and the function generator.