Giga-tronics 2500B Series Microwave Signal Generators
Performance Verification
Operation Manual, Part Number 34737, Rev C, June 30, 2011
Page 115
Amplitude Modulation Accuracy
Step
Action
11.
Set the AM Sensitivity of the UUT to 50 % and repeat steps 8 through 9 for each of the test frequen-
cies in Datasheet 6, recording the AM depth for each measurement in the appropriate column.
12.
Set the AM Sensitivity of the UUT to 75 % and repeat steps 9 and 10 for each of the test frequencies
in Datasheet 6, recording the AM depth for each measurement in the appropriate column.
13.
After you have entered all of the test results onto the datasheet, compare the test results with the
following to verify compliance of the 2500B:
End of Procedure
NOTE:
In the datasheet below, test the 2500B only at the frequencies within the range of your specific model of 2500B.
For frequencies not within your instrument’s range, enter “N/A” in the test results column(s) of the datasheet.
Table 72: Datasheet 6: Amplitude Modulation Accuracy Test
Datasheet 6: Amplitude Modulation Accuracy Test
Test
Frequency
Sensitivity
30 %/V
50 %/V
75 %/V
10 MHz
1
100 MHz
1
200 MHz
1
400 MHz
1
750 MHz
1
1500 MHz
1
3 GHz
6 GHz
12 GHz
26 GHz
36 GHz
40 GHz
Model type of 2500B series (for example, 2540B):
Serial number:
Tested by:
Date:
1