4. Programming Interfaces
Giga-tronics 2400/2500 Microwave Synthesizer Series
124
Programming Manual, Part Number 34783, Rev A, July 2009
Table 22 GT-12000 Native Commands: List Mode Commands
Number
Command Name
Comments
27
LGLF d HZ|KZ|KHZ|MZ|MHZ|GZ|GHZ
Set list range frequency to d.
28
LGI n
Insert frequency range as list n
29
LGIP n d
Insert power list range to the end of the
existing list. (Note: Parameter n and d are not
used, and kept in the command for backward
compatibility)
30
LR n
Set the existing list waiting on trigger to run.
(Note: Parameter n is not used, and kept in the
command for backward compatibility)
31
LS? n
Pre-compute list n
(Note: Parameter n is not used, and kept in the
command for backward compatibility)
32
RFB n
Set RF blanking off/on
(RFD 1 = RF blanking On, 0 = RF blanking OFF)
33
TR n
Set trigger mode (n=0 BNC, n=1 GPIB/GET)
34
SETYIGCAP n
Set YIG CAP in or out
If n = 0, YIG CAP is always out;
Else if n =1, YIG CAP switches with delay;
Else if n = 2, YIG CAP is in low noise.
35
YIGCAPDELAY n
Set YIG CAP delay in usec (100-2000 us)
36
YIGCAPDELAY?
Query YIG CAP delay time