Weighing parameters setting menu
Weight filter
This parameter is used to adjust the action of the digital filter applied to the detected
weight. If a low value is programmed, the action of the filter is lower while programming
a high value the weight is filtered more.
Filter value
Weight update frequency
0
25 Hz
1
25 Hz
2
25 Hz
3
25 Hz
4
12,5 Hz
5
12,5 Hz
6
12,5 Hz
7
6 Hz
8
6 Hz
9
6 Hz
Weight stability
Weight is considered stable when maintained within a certain weight range for a certain
period of time. This parameter determines the selection between different predefined
combinations indicated in the table below. In the presence of oscillations that tend to
change the weight of some units it is necessary to lower this value to be able to consider
it stable.
Stability value
Weight range
Time
0
Always stable
1
2 div
0,5 sec.
2
2 div
0,7 sec.
3
1 div
0,7 sec.
4
1 div
1 sec.
Filtr
StAb
Pagina –
page
–
Seite
27