GHM GREISINGER RS PRO 205-1444 Operating Manual Download Page 9

H61.0.RS.6C-04

 Operating Manual RS MH 3161-

… - ex 

page 9 of 14 

8  Output 

8.1  Serial Interface 

By means of the serial interface and a suitable electrically isolated interface adapter (USB 3100, USB 3100 N, 
GRS 3100 or GRS 3105) the device can be connected to a computer for data transfer.  
With the GRS 3105 up to 5 devices of the GMH3xxx- series can be connected to one interface (see also 
manual of GRS 3105). As a precondition the base addresses of all devices must not be identical, make sure  
to configure the base addresses accordingly 

(refer menu point “Adr.” in chapter 6). 

To avoid transmission errors, there are several security checks implemented e.g. CRC. 

The following standard software packages are available: 

 

GMHKonfig

Software for a comfortable editing o

f the device (e.g. Material selection…) 

 

EBS 20M / 60M

20-/60-channel software to display the measuring values 

In case you want to develop your own software we offer a 

development package

 including: 

a universally applicable Windows functions library ('GMH3000.DLL') with documentation that can be
used by the most programming languages. 

Suitable for Windows XP™, Windows Vista™, Windows 7™

Programming examples Visual Basic 4.0

™, Delphi 1.0™, Testpoint™

Note:  The measuring and display range values read back from the interface are always in the 

selected measurement unit (mbar, bar...)! 

Supported functions: 

Code  Name/Function 

Code  Name/Function 

Read measurement value 

200 

Read min display range 

Read system state 

201 

Read max display range 

Read min memory 

202 

Read display range - unit 

Read max memory 

204 

Read display range 

– decimal point 

12 

Read ID number 

208 

Read # of channels 

174 

Clear min memory 

214 

Read scale adjustment [%] 

175 

Clear max memory 

216 

Read offset adjustment 

176 

Read min measuring range 

222 

Read power off time (Conf-P.oFF) 

177 

Read max measuring range 

223 

Set power off time (Conf-P.oFF) 

178 

Read measuring range 

– measuring unit 

240 

Reset 

179 

Read measuring range 

– decimal point 

254 

Program version 

180 

Read kind of measuring of sensor 

199 

Read kind of measuring of display 

9  Input Adjustment 

9.1  Zero Displacement Sensor ('OFFS') 

A zero displacement can be carried out for the measured value: 

value displayed = value measured  -  offset 

Standard setting: 'off' = 0.0°, i.e. no zero displacement will be carried out. Together with the scale correction 
(see below) this factor is mainly used to compensate for sensor deviations. Input is in the display unit. 

9.2  Scale Correction Sensor ('SCAL') 

The scale of the measuring can be influenced by this setting (factor is in %): 

displayed value = measured value * (1+Scal/100) 

Standard setting: 'off' =0.000, i.e. value is not corrected. Together with the zero displacement (see above) 
this factor is mainly used to compensate for sensor deviations. 

Summary of Contents for GREISINGER RS PRO 205-1444

Page 1: ...H61 0 RS 6C 04 Operating Manual EN RS MH 3161 002 ex RS Stock number 205 1446 RS MH 3161 13 ex RS Stock number 205 1444 device without case device in leather case ready for use in Ex Zone rspro com ...

Page 2: ...3 CONNECTIONS 7 4 4 POP UP CLIP 7 5 START OPERATION 8 6 CONFIGURATION 8 7 REMARKS TO SPECIAL FEATURES 8 7 1 POWER OFF TIME 8 8 OUTPUT 9 8 1 SERIAL INTERFACE 9 9 INPUT ADJUSTMENT 9 9 1 ZERO DISPLACEMENT SENSOR OFFS 9 9 2 SCALE CORRECTION SENSOR SCAL 9 10 PRESSURE CONNECTION 10 10 1 DEVICE TYPE WITH RELATIVE PRESSURE 10 11 ERROR AND SYSTEM MESSAGES 10 12 SPECIFICATION 11 13 RESHIPMENT AND DISPOSAL 1...

Page 3: ...of the device misuse or malfunction of the connection or of the device The manufacturer is not liable for misprints 2 Safety 2 1 Intended Use The safety requirements see below have to be observed The device must be used only according to its intended purpose and under suitable conditions Use the device carefully and according to its technical data do not throw it strike it Protect the device from ...

Page 4: ...s pressure sensor power supply unit interface etc connected to the device must be on the same potentials If this is not guaranteed you have to connect them for a potential equalisation 7 Ambient conditions Take care that the device is not exposed to environments that make the intrusion of humidity water conducting liquids or dust possible 8 DANGER Trouble free operation and reliability of the devi...

Page 5: ... safe device again you have to check for visible damages and control functionality before you put it to the leather bag 14 Any changes or repair of the device is not allowed Please return device to manufacturer for repair or maintenance 3 Product Specification 3 1 Scope of supply The scope of supply includes Measuring device with 9V battery Operation manual Leather case 3 2 Operation and maintenan...

Page 6: ...figuration menu Store Quit press short hold function the last measuring value will be held in the secondary display press again hides the value Please Note Activating deactivating tara clears the max min memories 1 Zero Point Adjustment If there is no pressure or zero pressure absolute applied to the pressure ports the device will display 0 If there is a permanent deviation and device is operated ...

Page 7: ...dling Pull at label open in order to swing open the pop up clip Pull at label open again to swing open the pop up clip further Pop up clip closed Pop up clip at position 90 Pop up clip at position 180 Function The device with a closed pop up clip can be plainly laid onto a table or attached to a belt etc The device with pop up clip at position 90 can be set up on a table etc The device with pop up...

Page 8: ... 11 91 Base address of interface OFFS refer to list below The offset of sensor will be displaced by this value to compensate for deviations in the probe or in the measuring device oFF Zero displacement inactive 0 00 SCAL 2 000 2 000 The measuring scale of sensor will be changed by this factor to compensate deviations of temperature probe or measuring device oFF Scale correction factor inactive 0 0...

Page 9: ...interface are always in the selected measurement unit mbar bar Supported functions Code Name Function Code Name Function 0 Read measurement value 200 Read min display range 3 Read system state 201 Read max display range 6 Read min memory 202 Read display range unit 7 Read max memory 204 Read display range decimal point 12 Read ID number 208 Read of channels 174 Clear min memory 214 Read scale adju...

Page 10: ...000 2000 mbar 2000 0 mbar 11 Error And System Messages Display Meaning What to do Low battery power device will only continue operation for a short period of time Replace battery Battery empty Replace battery Mains operation without battery wrong voltage Check power supply replace it when necessary No display or confused characters device does not react on keypress Battery empty Replace battery Ma...

Page 11: ...or 8 x 1 mm 6 mm inner tube Ø tubes at the top of device Display 2 four digit LCDs 12 4 mm high and 7 mm high for measuring values and for min max memories hold function etc as well as additional functional arrows Pushbuttons 6 membrane keys Output 3 5 mm audio plug stereo Interface Serial interface 3 5mm jack can be connected to USB or RS232 interface of a PC via electrically isolated interface a...

Page 12: ...t package for reshipment especially for fully functional devices Please make sure that the device is protected in the package by enough packing materials 13 2 Disposal instructions Batteries must not be disposed in the regular domestic waste but at the designated collecting points The device must not be disposed in the unsorted municipal waste Send the device directly to us sufficiently stamped if...

Page 13: ...H61 0 RS 6C 04 Operating Manual RS MH 3161 ex page 13 of 14 16 EC Type Examination Certificate ...

Page 14: ...page 14 of 14 Operating Manual RS MH 3161 ex H61 0 RS 6C 04 17 EU Declaration of conformity ...

Reviews: