10 | Technical data
G 1114
24 / 29
B-H88.0.12.DB2-1.1
10
Technical data
Measuring range
0 .. 14000 hPa (mbar) abs.
0.00 .. 199.99 PSI abs.
0 .. 10500 mmHg (Torr) abs.
Accuracy
Typ.: ± 0.02 % FSS +/-0.1 % of value (@ 25 °C)
Max.: ± 0.1 % FSS ± 0.5 % of value
Measuring cycle
: approx. 25 measurements per second
: approx. 2.5 measurements per second
Overload
20000 hPa abs.
Pressure connection
1 hose connection, interchangeable with G1/8 universal port
Display
3-line segment LCD, additional symbols, illuminated (adjusta-
ble white, permanent illumination)
Standard function
Min/Max/Hold
Auto-power-Off function / if activated, switches the product off
automatically
Additional functions
: Tare function
V : Averaging over 2 s / 5 s / 10 s
Calibration
Zero point and gradient adjustment
Housing
Break-proof ABS housing
Protection rating
IP67
Dimensions L*W*H
[mm] and weight
108 * 54 * 28 mm without Pressure connector
140 g, incl. battery
Operating conditions
-20 to 50 °C; 0 to 95 % r.h. (short-term condensation possible)
Storage temperature
-20 to 70 °C
Current supply
2*AA battery (included in the scope of delivery)
Current requirement/
Battery life
approx. 0.6 mA (slow measurement SLO)
Operating time approx. 5000 h
Battery indicator
4-stage battery status indicator,
Charge indicator for low charge level: "BAT LO"
Auto-power-OFF function
The device switches off automatically if this is activated