G 1114
Operation | 5
B-H88.0.12.DB2-1.1
15 / 29
Parameter
Values
Meaning
Zero point correction
.
No offset
.
..
Offset active
Gradient compensation specification
.
No slope
.
- . .. .
Slope active
Sensor zero point correction offset
.
A zero point shift can be made for the measurement.
Displayed value = measured value offset
Standard setting: 0.0, i.e. no correction is made.
Sensor gradient correction
.
The gradient of the measurement can be influenced with this factor (factor is in %):
Displayed value = measured value * (1+Pr.SL/100)
Standard setting: 0.000, i.e. no correction is made.
The changed value is saved and the
is closed.
appears in the
display.
NOTE
If the product is switched off without saving the configuration, the last saved values
are reproduced on the next start-up of the product.