Frequently asked questions [FAQ]
Problem:
The logger cannot be stopped via keys.
Possible reason:
“Logger stop via buttons disabled” has been activated at the configu-
ration of the logger.
Remedy:
Deactivate this function by means of GSOFT 40K (at tab „settings“).
Problem:
The logger cannot be stopped both via keys and GSOFT 40K.
Possible reason:
“Do not stop active logger” has been activated at the configuration of
the logger.
Remedy:
Deactivate this function by means of EASYBus-configurator.
Problem:
The alarm settings or the sea-level correction cannot be
changed.
Possible reason:
The logger has still saved data.
Remedy:
Stop logger as recommended by the software (delete data).
Problem:
A error message is displayed.
Error message
Possible reason
procedure / remedy
measuring range
exceeded
Measuring has been ex-
ceeded.
Operate logger within its specifica-
tions.
Sensor defective
Send the logger to the manufac-
turer for repair
measuring range
under-run
Measuring range of this chan-
nel is under-run.
Operate logger within its specifica-
tions.
Sensor defective
Send the logger to the manufac-
turer for repair.
System error
System error
Reset the error via GSOFT 40K.
If the error message is displayed
furthermore send the logger to the
manufacturer for repair.
Sensorfehler
Sensor defective
Send the logger to the manufac-
turer for repair.
Berechnungsfehler
Sensor defective
Send the logger to the manufac-
turer for repair.