background image

E40.0.1X.6C-07 

Operating Manual EB-

Logg 80K… 

page 13 of 14 

11. 

Frequently asked questions [FAQ] 

Problem: 

The logger cannot be stopped via keys. 

Possible reason: 

“Logger stop via buttons disabled” has been activated at the configu-
ration of the logger. 

Remedy: 

Deactivate this function by means of GSOFT 40K (at tab „settings“). 

 

Problem: 

The logger cannot be stopped both via keys and GSOFT 40K. 

Possible reason: 

“Do not stop active logger” has been activated at the configuration of 
the logger. 

Remedy: 

Deactivate this function by means of EASYBus-configurator. 

Problem: 

The alarm settings or the sea-level correction cannot be 
changed. 

Possible reason: 

The logger has still saved data. 

 

Remedy: 

Stop logger as recommended by the software (delete data). 

Problem: 

A error message is displayed. 

Error message 

Possible reason 

procedure / remedy 

 

measuring range 
exceeded

 

Measuring has been ex-
ceeded. 

Operate logger within its specifica-
tions.  

Sensor defective 

Send the logger to the manufac-
turer for repair 

 

measuring range 
under-run 

Measuring range of this chan-
nel is under-run. 

Operate logger within its specifica-
tions. 

Sensor defective 

Send the logger to the manufac-
turer for repair. 

 

System error 

System error

 

Reset the error via GSOFT 40K. 
If the error message is displayed 
furthermore send the logger to the 
manufacturer for repair. 

 

Sensorfehler 

Sensor defective 

Send the logger to the manufac-
turer for repair. 

 

Berechnungsfehler 

Sensor defective 

Send the logger to the manufac-
turer for repair. 

 

 

 

Summary of Contents for GREISINGER EB log 80 K SMB2

Page 1: ...Y Tel 49 9402 9383 0 info greisinger de www greisinger de Installation and operating manual Data logger for temperature from version 1 0 EB Logg 80K Please read these instructions carefully before use...

Page 2: ...ASYBus Configurator 8 7 Battery life time and recording period 8 8 Operating displays 9 8 1 Measured value display 9 8 2 Logger status messages 9 8 3 System messages 10 8 4 Alarm and error messages 10...

Page 3: ...ufacturer for damages and consequential damages are voided with misuse disregard of this operating manual assignment of inadequately qualified technical personnel and arbitrary modifications of the pr...

Page 4: ...nly be guaranteed if the device is not subjected to any other climatic conditions than those stated under Specification 5 When connecting the device to other devices e g PC the interconnection has to...

Page 5: ...he logger s delivery status The delivered logger is in a sleep mode that means nothing is displayed and the power consumption is at its minimum When there is a communication with the software for the...

Page 6: ...ng attached on the device EB Logg 80KH sensor tube made of stainless steel 5 mm ap prox 50 mm long approx 1 m silicone cable with anti buckling glanding to housing EB Logg 80KH SMB1 sensor tube made o...

Page 7: ...n again Delete min value Push button for about 5 seconds CLr is displayed and the min values of all measuring channels are deleted 6 2 Operating possibilities at the device logger operation Function P...

Page 8: ...s and alarm delay 1 activate the function logger stop via buttons disabled 1 Please note The alarm settings and the sea level correction can be only changed if the logger is stopped logger without log...

Page 9: ...atus messages are dis played STOP The logger recording has been stopped The logger memory is empty Note At this state the power consumption is at its minimum HOLD The logger recording has been halted...

Page 10: ...end the logger to the manufacturer for repair 8 4 Alarm and error messages The following messages correspond to a particular measuring channel and are dis played accordingly ALARM LOW The measured val...

Page 11: ...ttery life time depending on the chosen measuring cycle about 5 years at 15 minutes and nominal temperature Measured value memory 250 000 measuring values of each variable Measuring cycle 4 seconds to...

Page 12: ...n AFK The sensor socket is ESD endangered when the sensor is disconnected Disconnecting the sensor in EASYBus mode is not permitted Maximum sensor cable length 3m 10 Decommissioning reshipment and dis...

Page 13: ...The logger has still saved data Remedy Stop logger as recommended by the software delete data Problem A error message is displayed Error message Possible reason procedure remedy measuring range exceed...

Page 14: ...ck view on pins EASYBus not used Connection examples a General connection diagram for a relay with self induction recuperation diode b Connection information for GR 10 GNR 10 or GNR 232 A GND to be co...

Reviews: