E40.0.1X.6C-07
Operating Manual EB-
Logg 80K…
page 10 of 14
8.3 System messages
BATTERY
:
• display of the BAT-arrow left hand down:
The battery is almost spent and needs to be replaced as soon as
possible.
The data recording is still active.
• permanent display
BAT
:
The battery is spent and the recording has been automatically
stopped (even if at EASYBus-operation).
It is still partly possible to read the measured values and to
read-out the saved logger data.
=> Please return the logger to the manufacturer.
ERROR
7
:
The logger has detected a system error.
• remedy: Reset the error via GSOFT 40K.
If the error message is displayed furthermore please
send the logger to the manufacturer for repair.
8.4 Alarm and error messages
The following messages correspond to a particular measuring channel and are dis-
played accordingly.
ALARM LOW
:
The measured value is below the selected min-alarm limit.
This and the current value is displayed alternately.
ALARM HIGH
:
The measured value is above the selected max-alarm limit.
This and the current value is displayed alternately.
ERROR
1
:
The measuring range of this channel is exceeded.
ERROR
2
:
The measuring range of this channel is under-run.
ERROR
9
:
The channel provides invalid values.
ERROR
11
:
The variable could not be calculated, because the measuring
range has been exceeded or a measured value, that is needed for
the calculation, is not available or faulty.
Note:
A description of the possible reasons for that error messages is in
chapter 12 (FAQ).