8.3  System messages 

 

 

BATTERY

• display of the  BAT-arrow left hand down: 

The battery is almost spent and needs to be replaced as soon as 
possible. 
The data recording is still active. 

• permanent display 

BAT

  The battery is spent and the recording has been automatically 

stopped (even if at EASYBus-operation).  

It is still partly possible to read the measured values and to  
read-out the saved logger data. 

 

=> Please return the logger to the manufacturer. 

 

ERROR 

7

The logger has detected a system error. 

• remedy:   Reset the error via GSOFT 40K. 

  If the error message is displayed furthermore please  
  send the logger to the manufacturer for repair. 

8.4  Alarm and error messages 

The following messages correspond to a particular measuring channel and are dis-
played accordingly. 

 

ALARM LOW

The measured value is below the selected min-alarm limit. 

This and the current value is displayed alternately. 

 

ALARM HIGH

The measured value is above the selected max-alarm limit.

 

This and the current value is displayed alternately.

 

 

ERROR 

1

The measuring range of this channel is exceeded. 

 

ERROR 

2

The measuring range of this channel is under-run. 

 

ERROR 

9

The channel provides invalid values. 

 

ERROR 

11

The variable could not be calculated, because the measuring  
range has been exceeded or a measured value, that is needed for 
the calculation, is not available or faulty. 
 

Note: 

A description of the possible reasons for that error messages is in 
chapter 12 (FAQ).  

 

 

Summary of Contents for GREISINGER EB log 80 K SMB2

Page 1: ...Y Tel 49 9402 9383 0 info greisinger de www greisinger de Installation and operating manual Data logger for temperature from version 1 0 EB Logg 80K Please read these instructions carefully before use...

Page 2: ...ASYBus Configurator 8 7 Battery life time and recording period 8 8 Operating displays 9 8 1 Measured value display 9 8 2 Logger status messages 9 8 3 System messages 10 8 4 Alarm and error messages 10...

Page 3: ...ufacturer for damages and consequential damages are voided with misuse disregard of this operating manual assignment of inadequately qualified technical personnel and arbitrary modifications of the pr...

Page 4: ...nly be guaranteed if the device is not subjected to any other climatic conditions than those stated under Specification 5 When connecting the device to other devices e g PC the interconnection has to...

Page 5: ...he logger s delivery status The delivered logger is in a sleep mode that means nothing is displayed and the power consumption is at its minimum When there is a communication with the software for the...

Page 6: ...ng attached on the device EB Logg 80KH sensor tube made of stainless steel 5 mm ap prox 50 mm long approx 1 m silicone cable with anti buckling glanding to housing EB Logg 80KH SMB1 sensor tube made o...

Page 7: ...n again Delete min value Push button for about 5 seconds CLr is displayed and the min values of all measuring channels are deleted 6 2 Operating possibilities at the device logger operation Function P...

Page 8: ...s and alarm delay 1 activate the function logger stop via buttons disabled 1 Please note The alarm settings and the sea level correction can be only changed if the logger is stopped logger without log...

Page 9: ...atus messages are dis played STOP The logger recording has been stopped The logger memory is empty Note At this state the power consumption is at its minimum HOLD The logger recording has been halted...

Page 10: ...end the logger to the manufacturer for repair 8 4 Alarm and error messages The following messages correspond to a particular measuring channel and are dis played accordingly ALARM LOW The measured val...

Page 11: ...ttery life time depending on the chosen measuring cycle about 5 years at 15 minutes and nominal temperature Measured value memory 250 000 measuring values of each variable Measuring cycle 4 seconds to...

Page 12: ...n AFK The sensor socket is ESD endangered when the sensor is disconnected Disconnecting the sensor in EASYBus mode is not permitted Maximum sensor cable length 3m 10 Decommissioning reshipment and dis...

Page 13: ...The logger has still saved data Remedy Stop logger as recommended by the software delete data Problem A error message is displayed Error message Possible reason procedure remedy measuring range exceed...

Page 14: ...ck view on pins EASYBus not used Connection examples a General connection diagram for a relay with self induction recuperation diode b Connection information for GR 10 GNR 10 or GNR 232 A GND to be co...

Reviews: