31
WARNING!
1.
Do not use
the option Smart-Thermo when your customer
➢
has got a pacemaker.
➢
is pregnant.
➢
is affected by thrombosis.
➢
suffers from pain due to unknown reasons.
2. Apply to healthy people
only.
3. When the treatment is stopped prematurely, cancel the Smart-
Thermo function by pressing the OFF-button on the control panel to
avoid heat accumulation.
4. Disconnect the MLE from the mains when it is not used for a longer
period of time, e.g. during the night.
WARNING!
Persons suffering from acute pain, venous disorders, low
blood pressure, inflammatory processes and other pathologies which
make heat-treatments or the positioning of the patient inappropriate
must
consult a qualified doctor before treatment.
WARNING!
Persons who are treated on the MLE (and / or are treated
with heat applications)
must not
be under the influence of intoxicating
substances (e.g. alcohol, drugs or medicine).
WARNING!
The cables of the models with Smart Thermo are between
the upholstery and metal rack (depending on model).
Removing of the
upholstery must be carried out by authorized professionals only!
11.1 Operating of the Smart Thermo heating system
❖