4-1
4.1 ELECTRIC POWERED STRETCHER
4
TROUBLE SHOOTING GUIDE
WARNING
To reduce the risk of electric shock, do not
remove the cover. Unit is to be serviced by
qualified service personnel (minimum 1 year
medical equipment service and repair experi-
ence) only.
If
Then
One motor or one col-
umn does not move but
all others are operating
correctly.
Step 1: Check all motor and column plug connections at the controller box.
Step 2: If a column does not move: Check the connection at the column.
Step 3: Perform the system reset procedure detailed in section 3.X.X
If the component does not run:
Contact
GF Health Products, Inc.
for further help and instruction.
Nothing moves
Step 1: Plug unit into a mains supply wall receptacle, observe the charge light
on the nurse control:
If the charge light is off:
Replace the controller
If the charge light is on:
1. Check the pendant control plug connection at the controller.
2. Ensure all lockout functions are deactivated.
3. Perform the system reset procedure detailed in section 3.X.X
4. Contact
GF Health Products, Inc.
for further help and instruction
The unit runs when
plugged into wall
receptacle, but does not
run on battery backup
Step 1: Plug unit into a wall receptacle overnight. If the batteries do not
hold a charge, replace the batteries in accordance with instructions in section
4.2.
Illustrated Part(s) breakdown is provided in Section 7 of this manual.
GF Health Products, Inc.
may be contacte
d
at
678.291.3
165
for additional information required to service or repair the equipment.
Always disconnect the power source when-
ever troubleshooting or servicing any elec-
tric powered stretcher.