Signet 8150 Flow Totalizer
Installation
Integral Installation with Field Mount Totalizer
3-8150-P0
3-8150-P1
3-8150-T0
3-8150-V0
(3-8510-XX)
(3-8051-X)
(3-8150-1)
Flow
Totalizer
Integral
Mounting Kit
Integral
Flow Sensor
Signet
Fittings
Integral Mount
•
Lubricate O-rings with a non-petroleum based, viscous
lubricant (grease) compatible with the system.
•
Using an alternating/twisting motion, lower the sensor
into the fitting, making sure the installation arrows on
the black cap are pointing in the direction of flow.
•
Engage one thread of the sensor cap then turn the
sensor until the alignment tab is seated in the
fitting notch.
Hand tighten the sensor cap.
DO NOT use any tools on the sensor
cap or the cap threads and/or fi tting
fl ange threads will be damaged,
1. Insert the wires from the sensor through the yellow conduit
base and locking ring.
2. Insert the locking ring into the conduit base, aligning it so the
square tab is close to the conduit ports.
3. Insert the mounting screw into the conduit base so the head
of the screw presses down on the locking ring
when tightened.
4. Connect sensor wires to the terminal connections on the
integral totalizer or terminal cover.
5. Remove the plastic pull tabs protecting the batteries, this will
cause the totalizer unit to power up.
6. For remote assembly, connect output wires to terminal
cover output.
7. Route the output wires through the conduit port in the conduit
base. Use a liquid-tight connector or conduit connector to
prevent moisture from entering the assembly.
8. Place totalizer or cover onto conduit base and twist to
lock in place.
Conduit Base Assembly Detail
The Mounting Screw for the
Locking Ring goes into the Conduit
Base (not the Locking Ring) -- the
screw head rests on the lip of the
Locking Ring.
Do not use this hole.
Sensor wires
Locking
Ring
Conduit
Base
Conduit
Ports
Mounting
Screw
Conduit
Base Kit
3-8150
Flow Totalizer
Square Tab
sensor
bale
sensor
cap
tab
black
conduit
cap
notch
Hand tighten only!
Plastic Sensor Installation Tips