‡ SIGNET

 5900 Salinity Monitor

6.2  CALIBRATE Menu

Requirements

System calibration (WET CAL) is required for first-time system setup or periodic sensor verification.  System calibration can be performed
with a solution of known salinity, or by comparison to a refractometer.  Refer to the WET CAL procedure for calibration details.

Press and hold ENTER key for 2 seconds:

is displayed.

Press keys in sequence to enter menu:

is displayed.

Menu Settings A - J:
(Factory default displays shown in menu
column 1)

A.

Selects cell type and cell value:
•  Standard cells: 10.0, or 20.0.  Use
10.0 or 20.0 settings.
•  Custom (certified) cells: 00.0000 -
999999.

B.

Selects displayed units; should be
selected for conductivity display INFO
ONLY.

C. Sets PPM factor when PPM display

units are selected, 0.01 - 9.99.

D. Sets Min-

>

Max meter dial range

(factory installed dial, 0 - 100).
Contact factory for custom dial
configurations.  Does not effect 4 to
20 mA output; set seperately.

E.

Selects WET CAL procedure for first
time system setup or periodic system
recalibration.

Menu items F - I repeat for relay 2 setup.

F.

Sets relay operation mode Low or
High, and setpoint, 00.0000 - 999999.
units.

G. Sets relay hysteresis, 00.0000 -

999999. units.  Set to zero to disable
feature

H. Sets relay minimum and maximum

pulse setpoint, 00.0000 - 999999.
units.

I.

Sets relay pulse rate, 000 - 300
pulses/minute.

J.

Sets user defined setup date for
maintenance records.  This feature is
not an internal timer or calendar.

CALIBRATE: ----

Enter Key Code

CALIBRATE: XXXX

Enter Key Code

Relay 1: Low

10.0000 PPT >

Last CAL:

02-05-98

Relay 1 Rate

120 Pulses/min

Relay 1 Pulse

00.0000

30.0000>

Relay 1 Pulse

10.0000

30.0000>

Relay 1 Hys:

02.0000 PPT >

Relay 1 Hys:

04.0000 PPT >

WET CAL:

Press <ENTER>

Min

Max: PPT

00000.0

0080.00

>

Min

Max: PPT

0.0000

40.000 

>

PPM Factor:

2.00

Units:

uS mS PPM k

 M

Units:

uS mS PPM k

 M

Cell: Custom

10.020

Cell: Standard
.01 .1 1 10 20

PPM Factor:

1.95

quick 

press

To return to 
VIEW:

quick 

press

To restore original value:

Mode (top row)
Setpoint (bottom row) 

Relay Low or High Mode Selected

Relay Pulse Mode Selected

A.

B.

D.

E.

F.

G.

J.

ENTER

"SAVING"

displayed

Choose:

Change:

Save:

To J

To A

Cell: Standard

10    >

Min

Max: PPT

0.0000

40.000 

>

WET CAL:

      >

Relay 1: Low

20.00 PPT >

Mode (top row)
Divisor (bottom row)

H.

Relay 1 Pulse

00.0000

30.0000>

I.

Relay 1 Hys:
4.0000 PPT >

Relay 1: Low

20.00 PPT >

Last CAL:

01-01-98  >

C.

PPM Factor:

2.00   >

Mode (top row)
Divisor (bottom row)

Last CAL:

01-01-98

Units:

mS

Relay 1 Rate

120 Pulses/min>

Relay 1 Rate

025 Pulses/min

Summary of Contents for SIGNET 5900

Page 1: ...ered non isolated maximum loop impedance 350 with a 12 V instrument supply voltage 950 with a 24 V instrument supply voltage To isolate output and prevent ground loop problems 1 Use monitor device wit...

Page 2: ...ents or operating currents that exceed the above specifications 5 2 HIGH alarm mode The relay is active when the solution salinity PPT rises above the setpoint and is inactive when the solution salini...

Page 3: ...0 100 See dial kit for additional information 6 1 VIEW Menu Menu Displays A G Factory default displays shown in menu column 1 A Active display of salinity PPT and temperature in degrees celsius C or f...

Page 4: ...int 00 0000 999999 units G Sets relay hysteresis 00 0000 999999 units Set to zero to disable feature H Sets relay minimum and maximum pulse setpoint 00 0000 999999 units I Sets relay pulse rate 000 30...

Page 5: ...any test solution for calibration purposes The 2 step WET CAL process first allows for verification or calibration of temperature followed by verification or calibration of salinity PPT using a known...

Page 6: ...A Adjust 3 88 mA Low High ENTER SAVING displayed A B C D E F Contrast 3 Display Decimal Display Average Low Set 4 mA 0 0000 PPT Set 20 mA 40 000 PPT 4 mA Adjust 4 00 mA 4 mA Adjust 4 00 mA Display Dec...

Page 7: ...acts Max voltage rating 5 A 30 VDC 5 A 125 VAC or 3 A 250 VAC power factor 1 0 Hysteresis User adjustable Current output 4 to 20 mA non isolated internally powered fully adjustable and reversible Upda...

Page 8: ...40 0000 WET CAL Relay 1 Low 10 0000 PPT Relay 1 Hys 4 0000 Relay 1 Pulse 10 0000 30 0000 Relay 1 Rate 120 Pulses min Last CAL 01 01 98 Sensor type and cell constant Process Units Min Max meter dial r...

Page 9: ...age 0 5 Off 0 sec Low 4 sec High 8 sec 00 0000 999999 00 0000 999999 3 0 5 0 mA 19 21 mA Celsius or Fahrenheit 0 00 9 99 3 Low 4 sec 00 0000 PPT 40 0000 PPT 4 00 mA 20 00 mA C 2 00 A B C D E F G H I M...