Pg 3 of 12

‡ SIGNET 

5900 Salinity Monitor

6.  Menu Functions

To access either CALIBRATE or OPTIONS menus, press and hold
the ENTER key as illustrated (right):

Menus:
• VIEW menu

:  Displayed during standard operation.  The

operator can navigate through the menu by pressing UP or
DOWN arrow keys.

• CALIBRATE Menu

:  Contains all critical display setup and

output parameters.  A security code feature prevents
unauthorized tampering.  The operator is required to enter an
access code for menu access.  The same code also unlocks
OPTIONS menus.

• OPTIONS Menu:

  Contains setup and display features

accessed for minor display or output adjustments.

• Reversible Dials (right)

The 5900 includes a dial kit with 6 reversible dial faces and
units decals (factory installed dial: 0 - 100).  See dial kit for
additional information.

6.1  VIEW Menu

Menu Displays A - G:
(Factory default displays shown in menu
column 1)

A. Active display of salinity (PPT) and

temperature in degrees celsius (

°

C) or

fahrenheit (

°

F).

B. Display of conductivity and

temperature.

C. Loop output display shows the loop

current output level.

D. Range display shows the programmed

min and max meter dial range.

E. Relay 1 display shows the

programmed operation mode and
setpoint for relay 1.

F. Relay 2 display shows the

programmed operation mode and
setpoint for relay 2.

G. Last calibration shows a user defined

setup date for maintenance records.
This feature is not an internal timer or
calendar.

OPTIONS

CALIBRATE

VIEW

2s

5s

Press &
hold for 
access:

ENTER

ENTER

0

100

0

10

PPT

PPT

A.

B.

C.

D.

Choose:

Change:Save:

To G

40.0 PPT

+25.0 

°

C

22.0 mS/cm

+25.0 

°

C

Loop Output:

20.00 mA

Min

Max: PPT

0.0000

40.0000

E.

Relay 1: Low

20.0000 PPT

F.

Relay 2: High

40.0000 PPT

To A

G.

Last CAL:

01-01-99

Summary of Contents for SIGNET 5900

Page 1: ...ered non isolated maximum loop impedance 350 with a 12 V instrument supply voltage 950 with a 24 V instrument supply voltage To isolate output and prevent ground loop problems 1 Use monitor device wit...

Page 2: ...ents or operating currents that exceed the above specifications 5 2 HIGH alarm mode The relay is active when the solution salinity PPT rises above the setpoint and is inactive when the solution salini...

Page 3: ...0 100 See dial kit for additional information 6 1 VIEW Menu Menu Displays A G Factory default displays shown in menu column 1 A Active display of salinity PPT and temperature in degrees celsius C or f...

Page 4: ...int 00 0000 999999 units G Sets relay hysteresis 00 0000 999999 units Set to zero to disable feature H Sets relay minimum and maximum pulse setpoint 00 0000 999999 units I Sets relay pulse rate 000 30...

Page 5: ...any test solution for calibration purposes The 2 step WET CAL process first allows for verification or calibration of temperature followed by verification or calibration of salinity PPT using a known...

Page 6: ...A Adjust 3 88 mA Low High ENTER SAVING displayed A B C D E F Contrast 3 Display Decimal Display Average Low Set 4 mA 0 0000 PPT Set 20 mA 40 000 PPT 4 mA Adjust 4 00 mA 4 mA Adjust 4 00 mA Display Dec...

Page 7: ...acts Max voltage rating 5 A 30 VDC 5 A 125 VAC or 3 A 250 VAC power factor 1 0 Hysteresis User adjustable Current output 4 to 20 mA non isolated internally powered fully adjustable and reversible Upda...

Page 8: ...40 0000 WET CAL Relay 1 Low 10 0000 PPT Relay 1 Hys 4 0000 Relay 1 Pulse 10 0000 30 0000 Relay 1 Rate 120 Pulses min Last CAL 01 01 98 Sensor type and cell constant Process Units Min Max meter dial r...

Page 9: ...age 0 5 Off 0 sec Low 4 sec High 8 sec 00 0000 999999 00 0000 999999 3 0 5 0 mA 19 21 mA Celsius or Fahrenheit 0 00 9 99 3 Low 4 sec 00 0000 PPT 40 0000 PPT 4 00 mA 20 00 mA C 2 00 A B C D E F G H I M...

Page 10: ...fy sensor wiring B Verify instrument temperature input Remove Black and White sensor wires from rear Temp IN and Iso Gnd terminals then place a 1100 resistor across terminals Power instrument and veri...

Page 11: ...Pg 11 of 12 SIGNET 5900 Salinity Monitor...

Page 12: ...nue El Monte CA 91731 2882 U S A Tel 626 571 2770 Fax 626 573 2057 For Worldwide Sales and Service visit our website www gfsignet com Or call in the U S 800 854 4090 GEORGE FISCHER Piping Systems 3 59...

Reviews: