background image

SAFETY INSTRUCTIONS

1.  Depressurize and vent system prior to installation or removal.

2.  Con

!

 rm chemical compatibility before use.

3.  Do not exceed maximum temperature/pressure speci

!

 cations.

4.  Wear safety goggles or faceshield during installation/service.

5.  Do not alter product construction.

6.  If this equipment is used in a manner not speci

!

 ed by the manufacturer, the protection provided 

by the equipment may be impaired.

7.  This device is not approved for use or installation in hazardous locations.

Signet 2537 Paddlewheel Flowmeter

3-2537.090 

Rev G   9/12  English

English

Specifi cations

General

Wetted Materials

Model Suffi x Sensor Body

Rotor

Pin

O-ring

-P0, -P1

glass-

!

 lled PP

PVDF, Black;

optional ETFE 

w/ or w/o carbon 

!

 ber reinforced 

PTFE sleeve

Titanium

FPM

-T0

PVDF, Natural

PVDF, Natural; 

optional ETFE 

w/ or w/o carbon 

!

 ber reinforced 

PTFE sleeve

PVDF, 

Natural

FPM

Case:   

 

PBT, yellow

Inside Cover: 

 

Valox, black

Wiring Ports: 

 

½

 in. NPT threads; liquid-tight  

 

 

 

 

connector accepts cables 7 to

 

 

 

10 mm OD (0.275 in. to 0.394 in.)

Power Requirements

Multi:

with Dry-Contact Relay:  24 VDC nominal, ±10%, regulated

 

 

 

30 mA max. current

with Solid-State Relay: 

5 to 24 VDC nominal, ±10%, regulated

 

 

 

30 mA max. current

Digital (S

3

L): 

 

5.0 VDC min. to 6.5 VDC max.

 

 

 

30 mA max. current (1.5 mA nominal)

4 to 20 mA: 

 

400 mV max. ripple voltage

 

 

 

30 mA max. current

Reverse Polarity and Short Circuit Protected: Up to 40 V, 1 hour

Over-Voltage Protection:  > 40 VDC over 1 hour

Description

The 2537 Paddlewheel Flowmeter Series offer low 

#

 ow, low power and high resolution with various output options such as a Volumetric 

Pulse, Pulse Divider, Flow Switch, Digital (S

3

L), or 4 to 20 mA. This unit can be con

!

 gured on-site directly through the built-in user 

interface.

•  The 4 to 20 mA model provides a blind current loop output.

•  The Digital (S

3

L) model provides a Digital (S

3

L) output for use with the Signet 8900 Multi-Parameter Controller or 9900 Transmitter.

•  The Multi model uses a single relay (mechanical or solid state) and has three selectable operating modes:

•  Divider Mode scales the paddlewheel frequency down to accommodate low frequency input devices.

•  Total Mode outputs one pulse per a set volume of 

#

 uid.

•  Flow Switch Mode uses a single relay for Hi or Lo alarm operation.

A small LCD enables the 2537 to be programmed without any external equipment. During normal operation the display is not visible.

For earlier versions of this sensor, please visit our manual archives at www.gfsignet.com and download the 2537 rev C manual.

Paddlewheel Sensor Performance Specifi cations

Pipe Size Range:  

DN15 to DN200 (

½

 in. to 8 in.)

Min. Reynolds Number:  4500

Paddlewheel Frequency:  49 Hz per m/s nominal

 

 

 

(15 Hz per ft/s nominal)

Operating Range:  

0.1 m/s to 6 m/s (0.3 ft/s to 20 ft/s)

Linearity: 

 

±1% of max. range @ 25 °C (77 °F)

Repeatability: 

 

±0.5% of max. range @ 25 °C (77 °F)

Electronics Performance Specifi cations

Input Frequency Range:  1 to 1000 Hz

System Response: 

100 ms update rate nominal

Environmental Requirements

Enclosure Rating:  

NEMA 4X/IP65

Storage Temperature: 

-10 to 75 °C (14 to 167 °F)

Ambient Temperature: 

0 to 65 °C (32 to 150 °F)

Relative Humidity: 

0 to 90% RH, non-condensing

Altitude:  

 

2000 m (6,562 ft)

Pollution Degree:  

2

Output Specifi cations

Signal Averaging:  

Programmable 0 to 100 seconds

Sensitivity Response: 

Programmable 0 to 9 scale

Pulse Divider/Total Pulse Output

Pulse Divider Setting:  

1.0000 to 99999

 

 

 

Maximum pulse rate: 300 Hz

 

 

 

Maximum pulse width: 50 ms

*3-2537.090*

Summary of Contents for Signet 2537

Page 1: ...Digital S3 L or 4 to 20 mA This unit can be con gured on site directly through the built in user interface The 4 to 20 mA model provides a blind current loop output The Digital S3 L model provides a...

Page 2: ...re and concentration Utilizing sensors in applications with aggressive substances is possible On special request GF Signet can provide sensors with a different internal resin encapsulation potting tha...

Page 3: ...when the system is in operation These con gurations are not recommended because it is dif cult to keep the pipe full Installation Pipe fittings Georg Fischer Signet offers a wide selection of install...

Page 4: ...irectly into the conduit port For conduit with ISO threads use the black thread adapter included with the connector kit To comply with NEC requirements do not use any metal conduit in the installation...

Page 5: ...e Dry Contact Relay requires an external power source of 24 VDC 10 The Solid State Relay requires an external power source of 5 to 24 VDC Solid State Relay requires a pull up resistor 10K ohm recommen...

Page 6: ...key S1 is held down for three seconds the ow rate is displayed for 10 minutes before reverting back to the model name In the Multi Model if the Multi menu item is set to divider then the divided puls...

Page 7: ...0 to 9 page 9 Mode Select the output mode Total or Divider or On OFF page 10 P Factor As PULSE DIVIDER divides input freq 1 0000 to 99999 page 10 P Factor As PULSE TOTAL multiplies K Factor 1 0000 to...

Page 8: ...tion Save the new setting Select the volumetric units for the ow measurement Flow Units Example Set the Flow Units to Gallons per minute This function is available on these versions of the 2537 2s 2s...

Page 9: ...ING only AVERAGING and SENSITIVITY Averaging and Sensitivity Settings Because ideal ow conditions are often impossible to achieve the ow rate is often erratic which causes erratic readings in control...

Page 10: ...Factor MULTIPLIED by the P factor The resulting frequency is output through a relay 2s Go to next menu item Return to Normal Operation Save the new setting Select DIVIDER or TOTAL operation The facto...

Page 11: ...hes the setpoint the 2537 will wait this long in seconds before triggering the alarm The factory setting is 0000 0 seconds 4 Program the DELAY A HI alarm will be activated when the ow rate rises ABOVE...

Page 12: ...T020 54 468 14 391 STAINLESS STEEL TEES ON SCH 40 PIPE 1 2 CR4T005 734 20 193 98 3 4 CR4T007 412 10 108 88 1 CR4T010 252 70 66 764 1 1 4 CR4T012 128 12 33 849 1 1 2 CR4T015 77 320 20 428 2 CR4T020 45...

Page 13: ...00 1 9868 8 CR4W080 4 3400 1 1466 CARBON STEEL WELDOLETS ON SCH 40 PIPE 2 1 2 CS4W025 37 600 9 9339 3 CS4W030 24 340 6 4306 4 CS4W040 13 920 3 6777 5 CS4W050 10 860 2 8692 6 CS4W060 7 5200 1 9868 8 CS...

Page 14: ...it breaks the sensor cannot be repaired 3 Install the new rotor by inserting one tip of the pin into the hole then carefully ex the opposite ear back enough to slip rotor into place H Dimensions The p...

Page 15: ...15 Signet 2537 Flowmeter Notes...

Page 16: ...Digital S3 L Integral Mount 3 2537 6C T0 159 001 320 4 to 20 mA Integral Mount Accessories Mfr Part No Code Description 3 2536 320 1 198 820 052 Rotor PVDF Black 3 2536 320 2 159 000 272 Rotor PVDF Na...

Reviews: