background image

3

GF 2751-7 pH Electronics

Lubricate O-rings with a non-petroleum based, viscous lubricant (grease) compatible with the system. 

It is not necessary to lubricate the electrodes body O-ring when using the 463X Flow Cell.

Do Not Use Lubricant or Sealing Tape on Threads.

Do Not Overtighten.

3610 Flow Tee Installation

1.  Lubricate the Dry-Loc and the electrode body O-rings 

with a non-petroleum lubricant.

2. 

Insert fully into the fitting.

3. 

Install bottom retaining nut onto the fitting and hand 

tighten. Do not use tools.

1.  Lubricate the Dry-Loc O-ring with a non-petroleum 

lubricant. (DO NOT lubricant the electrode body O-ring.) 

2.  Insert the sensor into the 2751-7.
3. 

Insert into the Chlorine (CI) flow chamber and push 

lightly to secure.

pipe tee

pipe tee

pipe tee

9950-X

CH 2

Installation

The bottom yellow nut is

not used when using the 2751-7 

pH electronics with the 463X Flow 

Cell.

3-2724-00 

Flat pH Electrode

463X Flow Cell

463X Flow Cell Installation

CHEMICAL COMPATIBILITY WARNING

The retaining nuts of pH and ORP sensors are not designed for prolonged contact with aggressive substances. 

Strong acids, caustic substances and solvents or their vapor may lead to failure of the retaining nut, ejection of 

the sensor and loss of the process fluid with possibly serious consequences, such as damage to equipment and 

serious personal injury. Retaining nuts that may have been in contact with such substances, e.g. due to leakage 

or spilling, must be replaced.

Wiring

Wiring to the 9950-X Chlorine Controller

To replace the 2751-7 pH electronics in a 463X Chlorine panel: 

• 

Measuring from the yellow housing threaded end, cut the cable to approximately 838 mm (33 in.).

• 

Remove approximately 10 mm (0.4 in.) of insulation and tin each conductor before inserting into terminals.

•  For calibration and configuration please refer to the 463X Chlorine Analyzer System and 

 

9950-X Chlorine Controller manuals.

CH1

+V
FREQ
DATA
GND

Black

White

X

No connection

Shield

Red

IN

OUT

pH

Cl

Reviews: