background image

 

Contamination of the sterilised goods  
“Danger of patient infection” 

See Section 2.4, page 4, item 2. 

)

 Important 

 

Only seal the packaging material when the red light 
is flashing, ie when the sealing temperature has 
been reached. 

 

If you reduce the temperature, the red light goes out 
(temperature too high). Wait until the temperature 
adjusts to the new setting and the red light starts to 
flash again. 

 

If you raise the temperature, the red light stays on 
continuously (without flashing) (temperature too low). 
Wait until the temperature adjusts to the new setting 
and the red light starts to flash again. 

5.3 Setting 

temperature 

level 10 internally 

The adjustment for temperature level 10 is on the underside of the case (1). 
This is how to set the temperature: 

− 

Switch off the sealer and unplug it from the power supply. 

 

GS-43-GB-002.doc  

7

Summary of Contents for GS 43

Page 1: ...er 4 2 5 Contamination of the sterilised goods Danger of patient infection 4 3 Installation 4 4 Overview of controls 5 5 Operating instructions 6 5 1 Switching on the sealer 6 5 2 Setting temperature...

Page 2: ...tted except by express consent Infringements are punishable and will lead to claims for damages We reserve all rights to exercise the provisions of commercial copyright law Warranty and liability We a...

Page 3: ...ll be accepted for use of the sealer for any other purpose The sealer may only be used for its intended purpose in accordance with the generally recognised rules of technology and the industrial healt...

Page 4: ...ackage will be difficult to peel open Paper fibres may be released and contaminate the sterile goods Released paper fibres may result in patient infection 3 Installation Set up the GS 43 sealer on a c...

Page 5: ...r guide to release the feed stop Move the feed stop to the required position and tighten the knurled nut With the feed stop fully forward the seam is about 5 mm away from the cut off edge of the mater...

Page 6: ...o flash The sealing temperature has been reached and the sealer is ready to seal 5 2 Setting temperature level 1 9 You can set the temperature level in ten steps Turn the switch A to the required posi...

Page 7: ...e temperature adjusts to the new setting and the red light starts to flash again If you raise the temperature the red light stays on continuously without flashing temperature too low Wait until the te...

Page 8: ...pulling it upwards as shown by the arrow Using a screwdriver 3 turn screw 4 clockwise to raise the temperature turn screw 4 anticlockwise to lower the temperature Refit the case cover 1 Danger of elec...

Page 9: ...ring breaks 6 Technical data SEALER GS 43 Dimensions w x d x h 570 x 280 x 240 mm approx Dimensions with paper guide 575 x 320 x 240 mm approx Weight 13 kg approx max sealing temperature 200 C Transpo...

Page 10: ...6 Pressure roller lower 4425723 Transport belt one 4416004 Toothed belts for pressure roller 4416016 Bearing bush 10 x 12 4415015 Ball bearing 4415000 From ____________________________________________...

Reviews: