31
Drain requirement
Flow capacity
100 L/min, 26.4 US Gal/min
Position (floor, Wall)
either
Drain connection size
Ø40 mm (1 9/16”) (Inner Dimension)
Ø46 mm (1 13/16”) (Outer dimension)
The bath must be connected to a drain. Use an air gap bracket, if needed.
Connections must be done in accordance with local regulations.
Physical chemical demands
PH range
Ph 7.5 - Ph 9
Conductivity (min)
More than 50 µS/cm
Chloride (max)
Less than 50 mg/L
Hardness
Less than 20° dH
Alkalinity
60-200 mg/l HCO
3
Electrical Data: EU and North America
Voltage
230V (EU)
120V (North America)
Frequency
50Hz (EU)
60Hz (North America)
Power consumption (max)
800VA
Number of phases
Single phase AC
Protective earth
Yes
Potential equalisation
Yes
Max Duty Cycle Lift / Hi-Lo
1 min ON, 9 min OFF
Max Duty Cycle Air Spa
15 min ON, 15 min OFF