18
Drain Installation
Drain installation - Floor
(Figure 1)
1. Connect the drainage hose to the pipe con-
nector and tighten with hose clamp.
2. Secure the open end of the pipe connector in
the water trap.
3. Attach the drainage hose to the floor
drain.
•
Install air gap bracket on the drain
(Air gap bracket can be ordered separately.)
(Figure 2)
After connecting the drain hose to the water trap
according to step 1 and 2 above,
1. Connect the open end of the drainage hose
to the pipe connector and tighten with hose
clamp.
2. Secure the open end of the pipe connector in
the elbow.
3. Cut the pipe adapter directly after the
threads.
(See Figure 2A.)
4. Apply the o-ring to the pipe adapter. Then
insert the adapter to the open end of the
elbow.
5. Now insert the open end of the pipe adapter to
the air gap bracket and tighten underneath
with the nut.
6. Attach the air gap bracket to the floor drain
by either of the following methods:
• lift the floor drain cover and fixate the
bracket using the M5 screws, M5 washers
and M5 Nyloc nuts,
or
• mark the floor drain cover through the
slots in the bracket and drill 2 x 3.5 mm
holes.
Secure the bracket in place using the self
tapping screws and M5 washers.
7. Ensure that the bath is situated so that it can
be operated without damaging the hose.
8. If the bath has been fitted with a
Multi-
Clean™
unit, the air gap bracket also has a
hole for the
MultiClean
hose to be con-
nected.
•
Drain requirement
Min. drainage capacity
100 l/min (26.4 US gal./min)
Drain connection size
Ø38 (1 1/2”) / Ø40 mm (1 9/16”) (Inner dimension)
Ø46 mm (1 13/16”) (Outer dimension)
The bath must be connected to a drain. Use an air gap bracket, if needed.
Connections must be done in accordance with local regulations.
NOTE
The drain installed to the floor or wall trap must comply with all local codes.
NOTE
See
MultiClean Assembly and
Installation Instructions
for details.
Summary of Contents for Arjohuntleigh Parker
Page 1: ...Parker Assembly and Installation Instructions with people in mind 06 AL 00_12GB 02 2016...
Page 4: ...4 Intentionally left blank...
Page 13: ...13 Intentionally left blank...
Page 23: ...23 HOT COLD Red Blue 1 2...
Page 32: ...32 Intentionally left blank...
Page 33: ...33 Intentionally left blank...
Page 34: ...34 Intentionally left blank...