Description
41
Measure the time it takes for an alarm to be reported and the time
needed to reach the patient. No more than ten seconds must pass
between these times. Observe the operating instructions for the
nurse call unit.
4.3
Membrane key panel
Do not apply excess pressure to the keys.
VitaGuard® recognizes key presses only when the keys have been
pressed for about one second.
There are six membrane keys on the top side of VitaGuard®.
Fig. 17 Keys on the top side