15
To charge the battery of the Com Box proceed as
follows:
Unplug the connecting cable of the measuring
probe from the jack (17) in the Com Box.
Align the marking on the plug of the USB power
cable ("charger") (18) with the marking on the
jack (17).
Insert the plug into the jack until it hits a stop.
Connect the USB power cable ("charger") with
the power supply unit to the power supply.
Connecting components
Take the equipment out of the transport
packaging.
Check the equipment for transport damage.
Contact the manufacturer if you detect any kind
of shipping damage.
DANGER
Sparks generated when connecting the
equipment can cause an explosion.
Connect the Com box and the
measuring probe always outside the
explosion-risk area!
To connect the Com box with the data collector
proceed as follows:
Align the marking on the plug (18) with the
marking on the jack (17).
Insert the plug into the jack until it hits a stop.
The back of the Com box is provided with
belt straps. If you do not want to carry the
Com box during the test procedure you
can attach it to your belt or to the supplied
strap.