13
to do so in the menu of the data collector once the
test has been finished.
For thermodynamic steam traps a certain limit
frequency of operations is used to evaluate its
performance.
If you did not enter the service pressure,
the equipment will only check whether the
temperature exceeds 40 °C. In this case
banking-up of condensate will not be
indicated.
For more information on how the readings are
evaluated and interpreted see the operating
instructions for the PC software program.
Storing and transporting the
equipment
Use only the supplied hard case for storing the
equipment.
If you do not use the equipment recharge the
data collector and the Com box at least every six
months.
Storing the equipment
Please observe the following items when storing
the equipment:
Protect the equipment and all components
against hard shocks and impacts.
Store the equipment only indoors.
The requirements specified under "Technical
Data" on page 31 must be observed.
Make sure that all these requirements are
always met when storing the equipment.
Please contact the manufacturer if you cannot
comply with the recommended storage
conditions.
Transporting the equipment
Meet the requirements for storage also when
transporting the equipment.
When transporting the equipment make sure
that you meet the requirements specified in the
section "Technical Data" on page 31.
Use only the supplied hard case for transporting
the equipment outside explosion-risk areas.
When transporting the equipment on site make
sure that it is not damaged by mechanical
shocks and cannot fall down.
The equipment must be connected and inside
the carrying bag if you want to transport the
equipment in explosion-risk areas.
When transporting the equipment observe
international regulations for the transport of
lithium-ion batteries.
Operation
Preparing tests
Recharging the batteries
DANGER
Sparks generated when charging the
equipment can cause an explosion.
Recharge the batteries of the Com box
and the data collector only outside the
explosion-risk area.
DANGER
Risk of explosion if a wrong USB power
cable ("charger") is used!
Use only the supplied USB power
cable ("charger") for recharging the
battery of the Com Box type
VKPC 40plus Ex