26
NRGT 26-2n - Installation & Operating Manual -
850879-00
Installation
Installing two level electrodes in a flange by detaching a terminal box
8.
Place the terminal box back on the electrode with the required orientation.
Ensure the terminal box is correctly orientated/aligned – position it correctly in advance.
9.
Tighten the fixing nut in the terminal box to a torque of 25 Nm.
10.
Reconnect the electrode wiring to the PCB, see table in
Fig 8
.
If necessary, join the connecting cables in the terminal box together using cable ties.
11.
Finally, check the wiring one more time.
12.
Close the rear panel of the second electrode terminal box and screw back on.