GERATHERM GT-5907 Instruction Manual Download Page 12

GB

38

c.  Once you choose the group, and 

if it already has memory, LCD will 

show the latest result with the LED 

indicator (picture 2). 

  If machine does not have memory 

before, the LCD will show six “0”.

 

d. One second later, LCD shows a 

single “0”. While the “down arrow” 
does not twinkle until the air pres-
sure is stable, and then it begins to measure automati-
cally.

e.  Then the monitor begins to pressurize automatically.

  During the measurement, the current pressure will dis-

play on the LCD, and the bars will automatically rise up 
or fall down with the current pressure in the cuff. The 

heart sign will glimmer with the heartbeat during the 

pressure are falling down.       

f.  After measurement, the result of 

systolic, diastolic and pulse rate 

will all display on LCD. And irregu-

lar heartbeat symbol and blood 

pressure classi

fi

 cation  indicator 

will twinkle (picture 3). The result 
will be automatically memorized by 

itself. And the COLOUR CHANGE 
LED will also indicate the blood 

pressure classi

fi

 cation.

g.  During the measurement, you can turn off the monitor 

by press the “START” button.

h.  Pressing “START” button after measuring, machine will 

be turn off, or it will be automatically shut down after 1 

minutes of none use.

Note

If the symbol “ 

 ” appears, it means the unit has 

detected irregular pulse during measurement. If the 
symbol appears regularly please consult a quali

fi

 ed 

physician for professional advice.

Operations

pic. 2

pic. 3

Summary of Contents for GT-5907

Page 1: ...GB Digital Blood Pressure Monitor for the upper arm INSTRUCTION MANUAL GT 5907...

Page 2: ...inches Designated use 28 Precautionary measures 29 Warranty 31 Information you should know before operating the unit 31 About the unit Function description 32 Explanation of displays 33 Preparation f...

Page 3: ...down comfortably with your elbows placed on the table and both feet on the ground Please do not interlock your legs during the measurement Wrap the cuff snug around your upper arm and lay your arm on...

Page 4: ...esult in circulatory disor ders which can cause unpleasant sensations such as lo calised bleeding under the skin or temporary numbness in your arm These symptoms do not usually last long How ever if y...

Page 5: ...ressure level What about low blood pressure In general a lower blood pressure reading is better unless causes unpleasant symptoms such as fainting and or light headedness The warranty for this blood p...

Page 6: ...red automatically and non invasively which helps to ease the operation The Geratherm med control have 2 x 99 memories with date and time display a big LCD screen and arrhythmia detection The RCC radio...

Page 7: ...to fit arm sizes between 22 and 30cm 8 7 to 11 81 inches Explanation of displays access for power supply cuff On Off Button LCD Display Memory Date Time Systo lic Pulse Diasto lic Low Batterie Memory...

Page 8: ...e type of alkaline bat teries be used to avoid incompatibility Rechargeable batteries with a voltage of 1 2V are not suitable for this product Keep batteries away from small children Do not throw batt...

Page 9: ...or use AC adapter accessory Use the device only with a medical approved stabi lized AC adapter 6V DC 600mA Note 1 No batteries are required during operation using the AC adapter 2 If AC adapter power...

Page 10: ...ata received the machine enters Clock Mode and keeps the time and date un changed When the machine is in the RCC Mode you can skip RCC receiving and enter the Clock Mode by pressing START button b For...

Page 11: ...l clock adjusting if no RCC signal or weak signal a When the machine is in Clock Mode press the START and MEM button at the same time abt 2 sec the machine enters Forced RCC Receiving mode then press...

Page 12: ...the heartbeat during the pressure are falling down f After measurement the result of systolic diastolic and pulse rate will all display on LCD And irregu lar heartbeat symbol and blood pressure class...

Page 13: ...ssment diastolic pressure mmHg systolic pressure mmHg Severe Hypertension Grade 3 Moderate Hypertension Grade 2 Mild Hypertension Grade 1 Optimal Normal High Normal Blood pressure classification Systo...

Page 14: ...ress the MEM but ton previously memories will show by turns d All memories have the WHO classification and the COL OUR CHANGE LED will also indicate the blood pres sure classification e If it has no m...

Page 15: ...o not drop the unit It is not shock proof Do not modify or disassem ble the unit or the arm cuff Do not twist the arm cuff Use a cloth moistened with water or neutral detergent to clean the body of th...

Page 16: ...not be used for a long time Do not press the START button if the cuff has not been properly wrapped around the arm Care and maintenance It is recommended the cuff should be disinfected 2 times every w...

Page 17: ...serious arrhyth mia to use this blood pressure monitor LCD shows low battery symbol Batteries low Change the batteries Observe the applicable regulations when disposing of the device and batteries Thi...

Page 18: ...of measurement the pressure reading must be below 15 mm Hg LCD shows Er 7 EEPROM accessing error LCD shows Er 8 Device parameter check error LCD shows Er A Span error BPM no response when you press bu...

Page 19: ...80 pulse beats min Accuracy 3mmHg blood pressure 5 pulse rate Memory 2 x 99 times with time and date displaying Low Battery Indicator yes Auto Power Off after 1 minute Battery Life Approx 200 measurem...

Page 20: ...ometers part 3 Supplementary requirements for electromechanical blood pressure measuring systems EN 1060 4 Non invasive sphygmomanometers Part 4 Test procedures to determine the overall system accu ra...

Page 21: ...requirements for EMC electromagnetic compatibility and defines the levels of immunity to electromagnetic interference and the maximum electromagnetic emission levels for medical devices This blood pr...

Page 22: ...umidity should be at least 30 Power frequency 50 60 Hz magnetic field IEC 61000 4 8 3 A m 3 A m Power frequency magnetic fields should be at levels character istic of a typical location in a typicalco...

Page 23: ...as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity ofequipment marked with the following symbol...

Reviews: