4.
Technical Data Diffustik
Modifications are not permitted with this device
Technical Data:
Dimensions:
Weight:
210 mm x 175mm x 75mm (L x w x h)
1120 g
Electrical data:
IP protection type:
Protection class:
Classification according to MDD:
Application component type:
PC interface:
Power supply:
Power consumption:
IPX0 IEC 529
II
IIa 93/42/EWG date 14.06.93 Appendix IX
BF according to VDE 0750 (DIN EN 60601-1)
USB 2.0
230 V AC
ca. 25 VA
Flow:
Flow sensor:
Measuring principle:
Measuring range:
Measuring range ventilation:
Resistance:
Effective dead space:
Flow resolution:
Sample rate:
Accuracy:
Ergoflow
Differential pressure
± 16l/s
0 – 300l/min
<0.12kPa/(l/s) <15l/s
<20ml
< 1ml/s
125 Hz
±3% or 20ml/s
Volume:
Measurement range:
Accuracy:
0 – 20L
± 3% or 50mL
CO Analyser:
Measurement principle:
Measurement range:
Accuracy:
Resolution:
Infra-red absorption
0 – 3000 ppm
< 1%
24 bit
CH4 Analyser:
Measurement principle:
Measurement range:
Accuracy:
Resolution:
Infra-red absorption
0 – 3000 ppm
< 1%
24 bit
Operating conditions:
Temperature:
Air humidity:
+15°C to +35°C
10 to 90% (non-condensing)
Storage and Transport
conditions:
Temperature:
Humidity:
Explosive conditions:
-10°C to +60°C
0 to 95% (non-condensing)
Device should not be used in explosive
atmospheres
Minimal requirements PC
system:
Norm:
Processor:
RAM Memory:
Hard disk:
Monitor:
Interface:
Operating system:
DIN EN 60950
Pentium III compatible or higher
1Ghz or higher recommended
500 MHz minimum
512 MB RAM or higher
192 MB RAM minimum
1.5 GB or greater recommended
600 MB minimum free space required
XGA (1024 x 768) or higher
USB 2.0 recommended
USB 1.1 minimum
Windows XP SP2 or higher