2.4

 

Maintenance 

 
The Diffustik has been designed to ensure it is a low maintenance 
device. Only minor maintenance is required to ensure trouble free 
operation.  

 

Interval 

Maintenance work 

Once a day 

Flow sensor, flow sensor housing and 
shutter unit cleaning and disinfect 

Once a day 

Check O-Rings for cracks/abrasion and 
change if necessary 

After 12 months 

Measurement check (Details can be 
found in the MTK protocol) 

 
 

 

Summary of Contents for Diffustik

Page 1: ...Diffustik User Manual Version 1 2 2 Date 17th October 2014...

Page 2: ...ON THE DIFFUSTIK 11 2 3 SECURITY 12 2 3 1 GENERAL SECURITY INFORMATION 12 2 3 1 1 NORMS AND GUIDELINES 12 2 3 1 2 ELECTROMAGNETIC COMPATIBILITY 13 2 3 2 SECURE USE OF DIFFUSTIK 13 2 3 3 CLEANING AND D...

Page 3: ...be integrated into even the largest of requirements This document contains important information for the operation of Blue Cherry and Diffustik We strongly recommend reading this manual carefully in o...

Page 4: ...ould only be connected to the electric power system of the cart The Software Blue Cherry is delivered with the Diffustik Its Standard version contains Measurements for SVC FVC CO Diffusion as well as...

Page 5: ...ic unit with USB connector Gas analyzer for CO and CH4 Housing with integrated electronic Blue Cherry Software CD Modular diagnostic platform operating under Microsoft Windows environment Quick User G...

Page 6: ...d disinfect the device The connectors of the Diffustik are located at the back side of the device Connect a grounded power supply cable to the AC 230V input 1 Now connect one of the bus connector 2 wi...

Page 7: ...urces which are connected to a reliable protective earth conductor 3 pin cable grounded and which specifications meet the details described on the label on the rear of the housing Extension cords must...

Page 8: ...t to the dovetail groove at the holder plate of the stand 3 After this plug the body of the electric unit 1 as far as it will go onto the closing head and connect the connection cable of the electric...

Page 9: ...cable to the jack of the electric unit 2 2 3 Installation of the demand valve If the Bodystik includes the option CO Diffusion the demand valve has to be connected additionally Therefore move the dema...

Page 10: ...connected accurately and without too much power The gas supply hose must only be connected to gas bottles with the following content up to 0 3 CH4 up to 0 3 CO up to 0 3 C2H2 optionally 21 O2 rest N2...

Page 11: ...Diffustik Connect the provided power supply to the jack of the Diffustik Now connect the Diffustik via the provided RS232 cable to a free RS232 port at your computer 2 2 6 Power on the Diffustik First...

Page 12: ...erstand this included user manual paying attention to the operating instructions Geratherm Respiratory is responsible for the security and reliability of function of the Diffustik as such only authori...

Page 13: ...hones close to the device There may be an influence to medical devices The interference resistance is 3V m Do not use a higher level close to the Diffustik 2 3 2 Secure use of Diffustik The Diffustik...

Page 14: ...Medical devices should not be used close to a patient always ensure a minimum distance of 1 5m between patient and non medical computer systems Leakage current may increase by connecting non medical...

Page 15: ...on the device don t put any objects on top of the device Never move foreign materials into the housing Never place the device on uneven or unsecured surfaces The device should not be used in dirty env...

Page 16: ...ensure all connections and cables are connected properly All additionally connected devices must comply with legal regulations and the total leakage current must not exceed the maximum allowed valued...

Page 17: ...hin patient area and resource B in medical used room Resource A within patient area and resource B in medical used room See remarks below table Resource A within patient area and resource B in a non m...

Page 18: ...ith local directives 2 3 6 Product labelling Diffustik Example of typical product labelling SN designates the serial number for the device The first 2 digits define the manufacturing year The example...

Page 19: ...designates the serial number for the device The first 2 digits define the manufacturing year The example 13 shows the device was manufactured in 2013 The identifier 8 402 is a product specific code Th...

Page 20: ...se Application component type BF according to DIN EN 60601 1 Indicates single use device Protection type II according to MDD Separate collection of electro and electronic devices IPX0 Indicates no pro...

Page 21: ...intenance device Only minor maintenance is required to ensure trouble free operation Interval Maintenance work Once a day Flow sensor flow sensor housing and shutter unit cleaning and disinfect Once a...

Page 22: ...be provided by Geratherm Respiratory or an authorized service partner of Geratherm Respiratory Geratherm Respiratory will not accept warranty claims on any unauthorised repairs 3 2 Warranty exclusion...

Page 23: ...5 Hz 3 or 20ml s Volume Measurement range Accuracy 0 20L 3 or 50mL CO Analyser Measurement principle Measurement range Accuracy Resolution Infra red absorption 0 3000 ppm 1 24 bit CH4 Analyser Measure...

Page 24: ...racy Resolution Infra red Absorption 0 3000 ppm 1 24 Bit CH4 Analyser Measurement principle Measurement range Accuracy Resolution Infrarot absorption 0 3000 ppm 1 24 bit Operating conditions Temperatu...

Reviews: