APPENDIX C: READING THE SENSOR
Reading the Turbo Sensor via Direct Read

 

While the Turbo comes with Seametrics’ easy to use Aqua4Plus software, you can also use 

standard Modbus

®

 RTU or SDI-12 equipment to easily take readings, so as to tie into your 

existing equipment or networks.
You may nee to use Aqua4Plus to set the baud rate. (You do not need to set the baud 

rate  for  SDI-12).    For  Modbus  you  must  have  Turbo  firmware  version  2.0  or  higher. 

 

For SDI-12 you must have version 2.2 or higher.

Reading Via Modbus

®

 Setting Baud Rate

Your  sensor  comes  configured  to 

communicate at 38,400 baud, with 8 data 

bits, one stop bit, and no parity. The sensor 

can also be set to 19,200 or 9600 baud, if 

needed for your application.
See Sensor Settings.

Taking Measurements

Read measurements using Modbus 

function 03 – Read Holding Registers.
Readings are located in two registers 

each, starting at address 62592. (AquiStar

®

 

register addressing is zero based, i.e., starts 

at zero. If your equipment uses one based 

addressing, you will need to add one to the 

register addresses.) 

Register addresses for Turbo Sensor

Zero Based

One Based

Temperature

62592

62593

Turbidity

62594

62595

Summary of Contents for Seametrics Turbo

Page 1: ...mpa n y ISO 9001 Precision Environmental Sensors An ONICON Brand Turbo Tubidity Smart Sensor with Data Logging For Sales Service Contact 2650 E 40th Ave Denver CO 80205 Phone 303 320 4764 Fax 303 322...

Page 2: ...TURBO INSTRUCTIONS Seametrics 253 872 0284 Page2 seametrics com 1997 2016 Seametrics All rights reserved Registered trademarks and trademarks belong to their respective owners...

Page 3: ...uter 5 Installing the Aqua4Plus Software 5 Installing the Sensor 5 Real time Data 6 Data Logging 6 Reports 8 Maintenance 11 Erratic Readings 11 Grounding Issues 11 Appendix A Technical Specifications...

Page 4: ...signs of damage on the appropriate shipping form After opening the carton look for concealed damage such as a cut cable If concealed damage is found immediately file a claim with the carrier Do s and...

Page 5: ...ing you to do the following Read via the Modbus protocol using your own software Read via SDI 12 protocol Display readings on a panel meter If you want to use one of these methods see Appendix C or co...

Page 6: ...Start readings default to table view To switch to Real time graphing view click the graph icon Real time readings default to a 1 second interval for 1 minute to adjust enter your desired settings her...

Page 7: ...on active under the Data Files tab as well as in the upper menu Once files have been started logged on the sensor they will be displayed under the Data Files tab Set Up Logging Window Here you will na...

Page 8: ...lder on your PC as well See Program Settings for default data folder location In the main view you ll see a list of reports sorted by date size or file name as selected here You can also search report...

Page 9: ...TION AND OPERATION Click to view data as a table Click to view data statistics Click to switch to full screen graphing view Graph saving and export options are available here You may change the displa...

Page 10: ...bidity is always in NTU The Information tab is a new feature allowing users to add metadata to their reports such as site location field notes or comments The Schedule tab will display the logging set...

Page 11: ...in the earth between the grounding points current can flow between the points creating an unexpected voltage difference and resulting erroneous readings The single most important step in minimizing a...

Page 12: ...feet Input Power 12 Vdc nominal 9 15 Vdc Ranges 0 400 NTU 0 3000 NTU For Modbus with 5 pin connector For SDI 12 with 5 pin connector For SDI 12 without connector White Purple Yellow Brown Blue Shield...

Page 13: ...e results Make sure the bottom face of the turbidity probe is submerged about 0 5 12mm into the calibration standard and at least 2 50mm from the bottom of the container to avoid interference with the...

Page 14: ...r stability while Aqua4Plus is measuring to ensure an accurate calibration To accept the second point reading click Ok Your new offset and slope values will appear to the right confirm other settings...

Page 15: ...or SDI 12 you must have version 2 2 or higher Reading Via Modbus Setting Baud Rate Your sensor comes configured to communicate at 38 400 baud with 8 data bits one stop bit and no parity The sensor can...

Page 16: ...aults to 48 or 12 seconds The recommended procedure to ensure proper warm up stabilization and to conserve power is Write Power On value of 48 48 4 12 seconds Power turns on Starts sampling about two...

Page 17: ...ss Query a a cr lf a cr lf Change Address a A b b cr lf change address from a to b Request measurement a M a 0022 cr lf request all measurements a D0 a 24 2412 458 73 cr lf read temperature and turbid...

Page 18: ...f request measurements a D0 a 458 73 cr lf read turbidity Concurrent measurement with CRC a CC a 0022 cr lf request all measurements w CRC a D0 a 24 2412 458 73 crc cr lf read temperature and turbidit...

Page 19: ...which cause damage to the products or cause abnormally frequent service calls or service problems h A new warranty period shall not be established for repaired or replaced material products or suppli...

Page 20: ...Seametrics 19026 72nd Avenue South Kent Washington 98032 USA P 253 872 0284 F 253 872 0285 1 800 975 8153 seametrics com LT 14418r10 20190117 1 17 2019...

Reviews: