Page 5
Operational Temp. Range
1
0 to 50° C (32 to 122° F)
Storage Temp. Range
-40 to 80° C (-40 to 176° F)
Calibrated Temp. Range
0 to 50°C (32 to 122° F)
Diameter
2.62 cm (1.03 in) OD
Length
14.43 cm (5.68 in)
Weight
170 g (0.37 lb)
Materials
Titanium body; Delrin® nose
cone, hanger, backend
Output Options
Rugged TROLL 100: USB or
RS232 via docking station
Rugged TROLL 200: USB or
RS232 via docking station;
Modbus/RS485 or SDI-12 via
Rugged TROLL 200 cable
Battery Type
3.6 V lithium internal battery
Battery Life
Typical 10 years or 2M
readings
1
External Power
Rugged TROLL 100: NA
Rugged TROLL 200: 8-36 VDC
Measurement Current
14 mA typical
Idle Current (network activity only) 200 µA typical
Sleep Current (no network activity) 20 µA typical
Memory
1.0 MB
Data Records
Date/time + 1 parameter
Date/time + 2 parameters
Date/time + 3 parameters
94,500
2
65,000
2
49,500
2
Number of logs supported
Rugged TROLL 100: 1 log
Rugged TROLL 200: 2 logs
Fastest Logging Rate
1 per second
Fastest Output Rate
Rugged TROLL 200 only
Modbus & SDI-12: 1 per second
Log Types
Linear, Fast Linear, Event
Suspension Cable (use with
backshell hanger & Rugged
TROLL 100/200 or Rugged
BaroTROLL)
Poly-coated 304 stainless steel
50 ft, 150 ft, 300 ft (15.24 m,
45.72 m, 91.44 m)
Rugged TROLL 200 Cable (use
with Rugged TROLL 200 &
Rugged BaroTROLL)
24/4 AWG
RS485 with connector,
RS485 stripped and tinned
Customizable up to 300 m
(1000 ft)
SDI-12 stripped and tinned
Standard lengths up to 60 m
(200 ft)
Rugged BaroTROLL
Range: 7.0 to 30.0 psi (0.5 to 2
bar, 0.5 to 2 atm, 14.25 to 61
in Hg)
Docking Station
(Rugged TROLL 100/200 &
Rugged BaroTROLL)
USB or RS232
Rugged TROLL Com
(use with Rugged TROLL 200 &
Rugged BaroTROLL)
USB or RS232 for RS485
communications via RS485
cable with connector
Minimum System Requirements
8JO4JUV WPSIJHIFS
400 MHz Intel® Pentium® II
processor; 128 MB RAM, 100 MB free disk space; Microsoft® Internet
Explorer® 6.01 or higher; Windows 2000 Professional SP4 or higher;
Windows XP Professional SP2 or higher; Windows Vista or Windows
7; CD-ROM drive; RS232 serial port or USB port
8JO4JUV.PCJMF
RuggedReader with Microsoft Windows Mobile®
operating system; at least 16 MB data storage memory (SD/CF
card or built-in non-volatile memory); Microsoft ActiveSync v. 4.5
or higher for Windows 2000/Windows XP. For Windows 7, download
and installation of the Windows Mobile Device Center software is
required for communication with the RuggedReader Handheld PC.
Pressure Sensor Type/Material
(Non-Vented)
Piezoresistive; Ceramic
Range
9.0 m (30 ft) (Burst: 60 ft)
30 m (100 ft) (Burst: 134 ft)
76 m (250 ft) (Burst: 368 ft)
Accuracy at 15° C
3
Typical ± 0.1% full scale (FS)
Accuracy (FS)
4
± 0.3% FS max.
Resolution
± 0.01% FS or better
Units of Measure
Pressure
Level
psi, kPa, bar, mbar, mmHg,
in, ft, mm, cm, m
Temperature Sensor
Silicon
Range
Calibrated: 32 to 122° F (0 to
50° C)
Accuracy
± 0.3° C
Resolution
0.01° C or better
Units of measure
Celsius or Fahrenheit
1
When used within the factory-calibrated temperature range
2
Data logged (no wrapping) from device within the factory-
calibrated temperature range
3
Across factory-calibrated pressure range
4
Across factory-calibrated pressure and temperature ranges
Specifications are subject to change without notice. Delrin is a
registered trademark of E.I. du Pont de Nemours & Co.
RUGGED TROLL 100 AND 200 SPECIFICATIONS