CORRECTING FOR BAROMETRIC PRESSURE

As shown in Fig. 3, the Rugged TROLL 100 and 200 instruments 

measure all pressure forces detected by their absolute pressure 

sensors. These forces include barometric (also called atmospheric) 

pressure and the pressure of the water above the sensor. As a 

result, logged data will show the effects of changes in barometric 

pressure. These effects can be removed manually (see the technical 

note “Manual Barometric Correction” in the Technical Resources 

section of the In-Situ website). To automatically remove barometric 

pressure effects, use Win-Situ Baro Merge software to post correct 

non-vented water level data. 

Using Baro Merge Software

You will need: 

r

Win-Situ Baro Merge and Win-Situ 5 installed on a desktop/

laptop PC

r

One or more Rugged TROLL logs in your Win-Situ working 

directory

r

A Rugged BaroTROLL log in your Win-Situ working directory 

covering the time period of the Rugged TROLL log(s) to be 

corrected

1.  Launch Baro Merge from the desktop shortcut or the 

Win-Situ 5 Tools menu.

2.  In the main window, select “Use a BaroTROLL file” for the 

compensation method (Fig. 4). 

TIP:

 Information on the other two compensation methods may be 

found in Baro Merge Help (press F1 in Baro Merge).

Fig 4. Baro Merge options with BaroTROLL file method selected

3.   Click 

  to select a Rugged BaroTROLL log file. (Baro 

Merge displays these automatically by site when they are in 

the proper location.) 

r

Click the plus sign to expand the site and display the logs. 

r

Check 

5

 the Rugged BaroTROLL file you want to use, then 

click OK 

 . 

4.  The main window returns to the screen with the Rugged 

BaroTROLL log file displayed (Fig. 4). Click Next 

 to 

continue.  

5.  Baro Merge reads and displays the values in the Rugged 

BaroTROLL log. Edit the values if necessary, or click Next 

 to continue. 

TIP:

 For details on adding or editing values and time stamps in the table, 

see Baro Merge Help (search for “Manual Entry”) .

6.   Now choose the Rugged TROLL log file(s) you intend to 

correct  (shown automatically when in the proper location). 

r

You may check 

5

 as many files as you like. For best results 

the time stamps in the Rugged BaroTROLL log and the 

Rugged TROLL log(s) should overlap. (Baro Merge will use 

the first and last barometric values for Rugged TROLL data 

points beyond the barometric data time stamps.)

7.   Click OK 

 . The barometric compensation is applied. 

When processing is complete, click OK again.

8.   Click 

  to close Baro Merge. 

r

 The compensated log files may be viewed or exported from 

the Data tab in Win-Situ. The original Rugged TROLL log file 

is not changed. A corrected log file with the same name and 

path is created with the file extension “–BaroMerge.wsl.” 

r

For more information, refer to the Win-Situ Baro Merge 

technical note at www.in-situ.com.

Warranty Provisions

In-Situ Inc. warrants the Rugged TROLL 100 and 200 and Rugged 

BaroTROLL for one (1) year, from the factory ship date, against defects 

in materials and workmanship under normal operating conditions. 

Complete warranty provisions are posted on our web site (www.
in-situ.com).

How to Contact Us

Technical Support: 

800 446 7488, option 3

(Toll-free 24 hours a day in the U.S. and Canada)
Mailing/Shipping Address: 

In-Situ Inc.

 

221 East Lincoln Ave.

 

Fort Collins, CO 80524

 

U.S.A.

Phone: 

970 498 1500

Fax: 

970 498 1598

Internet: 

www.in-situ.com

Email: 

[email protected]

To Obtain Repair Service (U.S.A.)

If you suspect that your Rugged TROLL 100/200 or Rugged 

BaroTROLL is malfunctioning and repair is re quired, you can help 

assure efficient servicing by following these guidelines: 

r

Call or email In-Situ Technical Support ([email protected]). 

Have the equipment with you when you call.

r

Be prepared to describe the problem, including how the 

instrument was being used and the conditions noted at the time 

of the malfunction. 

r

If Technical Support determines that service is needed, they 

will ask that you download and complete a Return Materials 

Authorization form available on the In-Situ website under 

Contact/Returns for Service.

r

Clean the instrument and/or cable. Decontaminate thoroughly if it 

has been used in a toxic or hazardous environment.  

r

Carefully pack the instrument and/or cable in a suitable shipping box. 

r

Mark the RMA number clearly on the outside of the box with a 

marker or label.

r

Send the package, shipping prepaid, to ATTN: RMA #XXXXX, at the 

address above.

The warranty does not cover damage during transit. We recom mend 

the customer insure all shipments. Warranty repairs will be shipped 
back prepaid.

Outside the U.S.A.

Contact your international In-Situ distributor for repair and service 

information.

Reviews: