7.3.6 START G-824A DATA TRANSMISSION
Now that the G-824A has been configured and the settings saved the settings in the
magnetometers flash memory, the magnetometer can be commanded to start outputting
readings from the Configuration menu by typing R. The following is observed:
Checking
The selected baud rate
Main RS-232 Baud Rate : 115200
is not a standard Baud rate.
<HR>
Press "Y" to ignore the warning.$$ 0.00000 26.7 62
$ 6338.32834 26.7 0
$ 20961.15197 26.7 0
$ 28898.15558 26.7 0
$ 28550.66295 3.5 0
$ 28550.66333 3.5 1
$ 28550.66359 3.5 0
7.3.7 Setting the Power Up Mode:
The G-824 can power up into two different modes. The default mode powers up and
enters the main configuration menu as follows:
File (Safe default values# (Not saved)
S/N 824113 CM321 configuration Version 00.02
Mag Code Version: 01.00 Build Date: 8/29/2013 @ 18:11:31
A Assign file name B Baud rates C Cycle time
D Display all settings E Erase FLASH F Format of output
G Gate control I Input voltages L Load from FLASH
M Mode of operation N Next (multi systems) O Other devices
P Password R Run magnetometer S Save parameters
T Triggering
From this menu you must choose “R” and press ENTER to start the magnetometer serial
output stream. The other power up mode goes directly into outputting data, skipping the
main configuration menu. To change between these modes, from the main configuration
menu chose “M” and press ENTER. You will see the following submenu listed below.
[Note: If the G-824 is configured to start up outputting data and skipping the
configuration menu then type a continuous stream of upper case Q’s until the main menu