Geometrics, Inc. G-822/G-823 Magnetometer Manual
27
A9997915900B3498000000
than 100,000 nT
).
A8677850800B3514000000
A7877821600B3645000000
A6997834700B3797000000
The same data in single channel Sandia format (ASCII):
A9977813100
A9989037600
A9995551700
A9999829300
A0007883500
A0003207100
A9997915900
A8677850800
A7877821600
A6997834700
Julian Time Set
The Set Time commands (D,H,M,S) will initialize the time in all counter modules. If a
particular counter has all Julian clock fields enabled the output string will have the
following inserted after the last A/D channel and before the CR/LF:
,DxxxHxxMxxSxx_xx
The x's would be ASCII characters (0-9) as required. The time registers are not incremented
until enabled with the 'J1xx' command, so they can be set up then synchronized by sending
the enable command at the correct time.
In Packed BCD and excess format the letters D,H,M,S, and _ are stripped and the data
encoded as per the Mag data above. The Day info is put into 2 bytes with the most
significant nibble of the most significant byte set to zero.
Julian Time Enable
This command starts/stops the Julian clock. To start the clock on counter 0 the command
would be "J100". "1" turns on the clock, while the "00" selects counter 0. To turn off the
clock update on counter 2 the command would be "J002".