5
Channel Number
Terminal Strip
Designator
Terminal Strip
Position
Description Cable
Wire
Color
12
T12
23H
Vibrating Wire +
RED
12 T12 23L Vibrating
Wire
-
BLACK
12 T12 24H Thermistor
+ GREEN
12 T12 24L Thermistor
- WHITE
12
T12
S12
Analog Ground (shield)
BARE WIRE
13
T13
25H
Vibrating Wire +
RED
13 T13 25L Vibrating
Wire
-
BLACK
13 T13 26H Thermistor
+ GREEN
13 T13 26L Thermistor
- WHITE
13
T13
S13
Analog Ground (shield)
BARE WIRE
14
T14
27H
Vibrating Wire +
RED
14 T14 27L Vibrating
Wire
-
BLACK
14 T14 28H Thermistor
+ GREEN
14 T14 28L Thermistor
- WHITE
14
T14
S14
Analog Ground (shield)
BARE WIRE
15
T15
29H
Vibrating Wire +
RED
15 T15 29L Vibrating
Wire
-
BLACK
15 T15 30H Thermistor
+ GREEN
15 T15 30L Thermistor
- WHITE
15
T15
S15
Analog Ground (shield)
BARE WIRE
16
T16
31H
Vibrating Wire +
RED
16 T16 31L Vibrating
Wire
-
BLACK
16 T16 32H Thermistor
+ GREEN
16 T16 32L Thermistor
- WHITE
16
T16
S16
Analog Ground (shield)
BARE WIRE
Table 1 - Transducer Cable Connections
Insert four D-cells straight down into the battery holders. Ensure the polarity of the batteries
matches the diagram on the battery holder. Note that there is a zip-lock bag containing four
desiccant packs shipped along with the batteries. As soon as the batteries are installed, take the
desiccant packs out of the zip-lock bag and place them inside the enclosure. Immediately close
and reseal the lid. This will help to prevent condensation of moisture within the enclosure. If
batteries and desiccants are installed in the office, be sure to seal the lid before going out to the
field.
2.2 Earth Ground Installation
The LC-2x16 provides lightning protection in the form of gas tube surge arrestors. In order for
these components to divert the energy from a lightning strike safely to ground, a good solid
electrical connection to earth ground needs to be made. A grounding rod should be driven (or
other suitable attachment to earth utilized) to ground the system and provide a path to earth in the
event of a lightning strike. A 6' to 8' copper stake with appropriate large gauge wire (12 AWG or
larger) connected to the LC-2x16 enclosure is suggested. The stake should be driven as close to
the datalogger as possible, and to a depth of at least three feet (one meter). A copper grounding
lug is supplied on the exterior of the LC-2x16 enclosure to provide connection to this wire from
the grounding rod.